Insulation discontinuity in a vagus nerve stimulator lead: a treatable cause of intolerable stimulation-related symptoms.
Spitz, Mark C; Winston, Ken R; Maa, Edward H; Ojemann, Steven G.
Afiliação
  • Spitz MC; Department of Neurology, The University of Colorado at Denver Health Science Center, Denver, Colorado, USA.
J Neurosurg ; 112(4): 829-31, 2010 Apr.
Article em En | MEDLINE | ID: mdl-19558306
ABSTRACT
Discontinuity in the silicone insulation over an electrode of a left vagus nerve stimulator (VNS) allowed the aberrant leak of current to the phrenic nerve and other structures. This resulted in ipsilateral diaphragmatic dysfunction, inability to vocalize, and severe radiating pain into the jaw and upper incisor for the duration of each stimulation. The device was explanted and a new device was implanted. All stimulation-related symptoms ceased immediately. A similar discontinuity in the silicone insulation is the likely explanation for several prior reports of poorly understood pains and phrenic nerve stimulation in patients with VNSs. The findings and analysis of this case establish a rationale for consideration of replacement of the VNS lead in all similarly symptomatic patients.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Silicones / Dor Facial / Eletrodos Implantados / Estimulação do Nervo Vago Tipo de estudo: Diagnostic_studies / Etiology_studies Limite: Female / Humans / Middle aged Idioma: En Revista: J Neurosurg Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Silicones / Dor Facial / Eletrodos Implantados / Estimulação do Nervo Vago Tipo de estudo: Diagnostic_studies / Etiology_studies Limite: Female / Humans / Middle aged Idioma: En Revista: J Neurosurg Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Estados Unidos