Perfusion MRI of brain tumours: a comparative study of pseudo-continuous arterial spin labelling and dynamic susceptibility contrast imaging.
Järnum, Hanna; Steffensen, Elena G; Knutsson, Linda; Fründ, Ernst-Torben; Simonsen, Carsten Wiberg; Lundbye-Christensen, Søren; Shankaranarayanan, Ajit; Alsop, David C; Jensen, Finn Taagehøj; Larsson, Elna-Marie.
Afiliação
  • Järnum H; Department of Radiology, Aalborg Hospital/Arhus University Hospital, Hobrovej 18-22, 9000, Aalborg, Denmark. h.jarnum@rn.dk
Neuroradiology ; 52(4): 307-17, 2010 Apr.
Article em En | MEDLINE | ID: mdl-19841916
INTRODUCTION: The purpose of this study was to compare the non-invasive 3D pseudo-continuous arterial spin labelling (PC ASL) technique with the clinically established dynamic susceptibility contrast perfusion magnetic resonance imaging (DSC-MRI) for evaluation of brain tumours. METHODS: A prospective study of 28 patients with contrast-enhancing brain tumours was performed at 3 T using DSC-MRI and PC ASL with whole-brain coverage. The visual qualitative evaluation of signal enhancement in tumour was scored from 0 to 3 (0 = no signal enhancement compared with white matter, 3 = pronounced signal enhancement with equal or higher signal intensity than in grey matter/basal ganglia). The extent of susceptibility artefacts in the tumour was scored from 0 to 2 (0 = no susceptibility artefacts and 2 = extensive susceptibility artefacts (maximum diameter > 2 cm)). A quantitative analysis was performed with normalised tumour blood flow values (ASL nTBF, DSC nTBF): mean value for region of interest (ROI) in an area with maximum signal enhancement/the mean value for ROIs in cerebellum. RESULTS: There was no difference in total visual score for signal enhancement between PC ASL and DSC relative cerebral blood flow (p = 0.12). ASL had a lower susceptibility-artefact score than DSC-MRI (p = 0.03). There was good correlation between DSC nTBF and ASL nTBF values with a correlation coefficient of 0.82. CONCLUSION: PC ASL is an alternative to DSC-MRI for the evaluation of perfusion in brain tumours. The method has fewer susceptibility artefacts than DSC-MRI and can be used in patients with renal failure because no contrast injection is needed.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Encéfalo / Neoplasias Encefálicas / Imageamento por Ressonância Magnética / Imagem de Perfusão Tipo de estudo: Observational_studies / Qualitative_research Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Neuroradiology Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Dinamarca País de publicação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Encéfalo / Neoplasias Encefálicas / Imageamento por Ressonância Magnética / Imagem de Perfusão Tipo de estudo: Observational_studies / Qualitative_research Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Neuroradiology Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Dinamarca País de publicação: Alemanha