Your browser doesn't support javascript.
loading
Atmospheric NO(2) determination by 442-nm laser induced fluorescence.
Appl Opt ; 14(6): 1418-22, 1975 Jun 01.
Article em En | MEDLINE | ID: mdl-20154840
ABSTRACT
An atmospheric monitor with a detectability of 0.6 ppbv, utilizing a compact He-Cd laser at 442 nm for excitation of the NO(2) molecules is described. Photon-counting techniques were employed to detect the NO(2) fluorescence transmitted through a bandpass liquid solution filter that did not fluoresce upon absorption of scattered laser light. Operational characteristics of the instrument are described. Results obtained in ambient air monitoring are presented.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Appl Opt Ano de publicação: 1975 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Appl Opt Ano de publicação: 1975 Tipo de documento: Article