Inadequate gas supply to patients with an adjustable pressure-limiting valve in the fully opened position.
Hirabayashi, Go; Uchino, Hiroyuki; Ogihara, Yukihiko; Omi, Akibumi; Ohmura, Akito.
Afiliação
  • Hirabayashi G; Department of Anesthesiology, Hachioji Medical Center, Tokyo Medical University, 1163 Tatemachi, Hachioji, Tokyo, 193-0998, Japan. goh@tokyo-med.ac.jp
J Anesth ; 24(4): 626-9, 2010 Aug.
Article em En | MEDLINE | ID: mdl-20464429
ABSTRACT
With spontaneous ventilation, sufficient exhaust resistance at the adjustable pressure-limiting (APL) valve when fully opened ensures that the reservoir bag fills and adequately supplies gas to patients. A lack of exhaust resistance with the APL valve fully open caused inadequate gas supply to patients with four types of anesthesia machines SA2 (Dräger), Excel-210 SE (GE), Fabius (Dräger), and Cato (Dräger). Mechanically, the SA2 and Excel-210 SE APL valve systems, which are of the spring-loaded disc type positioned horizontally, cannot maintain sufficient exhaust resistance with the APL valve fully open. As for the Fabius and Cato, an exhaust valve independent of the APL valve should maintain sufficient exhaust resistance continuously. However, accumulated viscous substances on the thin diaphragm of the exhaust valve contributed to hindrance of diaphragm closure.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ventiladores Mecânicos / Anestesiologia / Erros de Medicação Limite: Humans Idioma: En Revista: J Anesth Assunto da revista: ANESTESIOLOGIA Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ventiladores Mecânicos / Anestesiologia / Erros de Medicação Limite: Humans Idioma: En Revista: J Anesth Assunto da revista: ANESTESIOLOGIA Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Japão