Effects of ambient-light correction in luminance measurements of liquid-crystal display monitors by use of a telescopic-type luminance meter.
Radiol Phys Technol
; 3(1): 65-9, 2010 Jan.
Article
em En
| MEDLINE
| ID: mdl-20821104
The quality control of liquid-crystal display (LCD) monitors has become one of the important topics for maintaining reliable soft-copy readings in the interpretation of diagnostic images. In this paper, the effects of correction in the luminance measurement of an LCD monitor by use of a telescopic-type luminance meter were investigated. The luminance of the LCD monitor in different ambient-lighting conditions was measured and compared to the results obtained with no ambient lighting (0 lux). The reproducibility of luminance measurements and luminance ratios without a baffled tube was lower than those measured with the baffled tube due to the effect of ambient light. These tendencies were obvious at a relatively low luminance. The correction method by subtraction of the reflected ambient light on the surface of the LCD monitor and the stray light of the telescopic-type luminance meter from the measured luminance was examined. We found that the correction was able to bring the luminance close to that measured with the baffled tube.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Apresentação de Dados
/
Cristais Líquidos
/
Luz
Idioma:
En
Revista:
Radiol Phys Technol
Assunto da revista:
BIOFISICA
/
RADIOLOGIA
Ano de publicação:
2010
Tipo de documento:
Article
País de afiliação:
Japão
País de publicação:
Japão