ABSTRACT
Protective patient equipment for CT examinations is not routinely provided. The aim of this study was to determine whether, and if so what, specific protective equipment is beneficial during CT scans. The absorbed organ doses and the effective doses for thorax, abdomen/pelvis and brain CT investigation with and without the use of protective patient equipment have been determined and compared. All measurements were carried out on modern multislice CT scanner using an anthropomorphic phantom and thermoluminescence dosemeters. The measurements show that protective equipment reduces the dose within the scattered beam area. The highest organ dose reduction was found in organs that protrude from the trunk like the testes or the female breasts that can largely be covered by the protective equipment. The most reduction of the effective dose was found in the male abdomen/pelvis examination (0.32 mSv), followed by the brain (0.11 mSv) and the thorax (0.06 mSv). It is concluded that the use of protective equipment can reduce the applied dose to the patient.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Tomografia Computadorizada por Raios X / Contagem Corporal Total / Imagens de Fantasmas / Modelos Anatômicos / Modelos Biológicos Tipo de estudo: Health_economic_evaluation / Prognostic_studies Limite: Female / Humans / Male Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Suíça

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Tomografia Computadorizada por Raios X / Contagem Corporal Total / Imagens de Fantasmas / Modelos Anatômicos / Modelos Biológicos Tipo de estudo: Health_economic_evaluation / Prognostic_studies Limite: Female / Humans / Male Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Suíça