ABSTRACT
BACKGROUND:
We studied which factor could predict aseptic loosening in ABG I hip prosthesis with hydroxyapatite coating. Aseptic loosening and periprosthetic osteolysis are believed to be caused, at least in part, by increased polyethylene (PE) wear rate via particle disease. Based on it, increased PE wear rate should be associated with aseptic loosening regardless of the type of implant.METHODS:
We analyzed data from 155 revisions of ABG I hip prostheses to examine the influence of patient, implant, surgery, and wear related factors on the rate of aseptic loosening at the site of the cup. This was calculated by stepwise logistic regression analysis. The stability of the implant and severity of bone defects were evaluated intraoperatively.RESULTS:
We found that men (odds ratio, OR = 5.6; p = 0.004), patients with Charnley class C (OR = 6.71; p = 0.013), those having more severe acetabular bone defects (OR = 4 for each degree of severity; p = 0.002), and longer time to revision surgery (OR = 1.51 for each additional year; p = 0.012) had a greater chance of aseptic loosening of the cup. However, aseptic loosening was not directly predicted by polyethylene wear rate in our patients.CONCLUSION:
Severity of bone defects predicts the risk for aseptic loosening in ABG I cup. Factors potentially associated with the quality of bone bed and biomechanics of the hip might influence on the risk of aseptic loosening in this implant.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Complicações Pós-Operatórias
/
Falha de Prótese
/
Osteoartrite do Quadril
/
Artroplastia de Quadril
/
Articulação do Quadril
Tipo de estudo:
Diagnostic_studies
/
Etiology_studies
/
Prognostic_studies
/
Risk_factors_studies
Aspecto:
Determinantes_sociais_saude
Limite:
Aged
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
BMC Musculoskelet Disord
Assunto da revista:
FISIOLOGIA
/
ORTOPEDIA
Ano de publicação:
2010
Tipo de documento:
Article
País de afiliação:
República Tcheca