Quantification of direct current in electrically active implants using MRI methods.
Wojtczyk, Hanne; Graf, Hansjörg; Martirosian, Petros; Ballweg, Verena; Kraiger, Markus; Pintaske, Jörg; Schick, Fritz.
Afiliação
  • Wojtczyk H; Section on Experimental Radiology, Department of Diagnostic and Interventional Radiology, University Hospital Tübingen, Eberhard Karls University Tübingen, Hoppe-Seyler-Straße, Germany. hanne.wojtczyk@yahoo.de
Z Med Phys ; 21(2): 135-46, 2011 May.
Article em En | MEDLINE | ID: mdl-21277177
The aim of this study was to evaluate a variety of phase- and magnitude-based MRI methods at 1.5 T and 3 T regarding their sensitivity and accuracy with respect to the quantification of electrical direct current via the induced magnetic field inhomogeneity. For this, a phantom was constructed which was specially designed to reduce RF effects and which provided a one-dimensional electrical direct current in a thin copper conductor perpendicular to the static magnetic field of the scanner. The current was varied between 4 mA and 472 mA. The analysis of FLASH phase images as well as trueFISP and MAGSUS images revealed that the accuracy of the MR current measurement depended on the method and the field strength: the mean of the absolute deviations of the measured current values from the adjusted current values varied between 9% and 21%. The phase measurement with a FLASH sequence was found to be more sensitive than the trueFISP and MAGSUS measurements. In FLASH magnitude images as well as in images of spin echo sequences with on- and off-resonant frequency selective saturation pulses the extension of the artifact increased with the electrical current. MRI methods for the quantification of electrical direct current might e.g. play a role in functional testing of electrically active devices in the human body in terms of measuring the present current. One-dimensional electrical direct current in a thin, straight conductor could also be applied to the visualization of instruments in interventional MRI procedures. Currents below 100 mA would be sufficient to create distinct artifacts, at least under simplified conditions (homogeneous background etc.).
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Próteses e Implantes / Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética / Terapia por Estimulação Elétrica / Artefatos / Cobre / Condutividade Elétrica Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Z Med Phys Assunto da revista: RADIOTERAPIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Alemanha País de publicação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Próteses e Implantes / Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética / Terapia por Estimulação Elétrica / Artefatos / Cobre / Condutividade Elétrica Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Z Med Phys Assunto da revista: RADIOTERAPIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Alemanha País de publicação: Alemanha