Quantification of direct current in electrically active implants using MRI methods.
Z Med Phys
; 21(2): 135-46, 2011 May.
Article
em En
| MEDLINE
| ID: mdl-21277177
The aim of this study was to evaluate a variety of phase- and magnitude-based MRI methods at 1.5 T and 3 T regarding their sensitivity and accuracy with respect to the quantification of electrical direct current via the induced magnetic field inhomogeneity. For this, a phantom was constructed which was specially designed to reduce RF effects and which provided a one-dimensional electrical direct current in a thin copper conductor perpendicular to the static magnetic field of the scanner. The current was varied between 4 mA and 472 mA. The analysis of FLASH phase images as well as trueFISP and MAGSUS images revealed that the accuracy of the MR current measurement depended on the method and the field strength: the mean of the absolute deviations of the measured current values from the adjusted current values varied between 9% and 21%. The phase measurement with a FLASH sequence was found to be more sensitive than the trueFISP and MAGSUS measurements. In FLASH magnitude images as well as in images of spin echo sequences with on- and off-resonant frequency selective saturation pulses the extension of the artifact increased with the electrical current. MRI methods for the quantification of electrical direct current might e.g. play a role in functional testing of electrically active devices in the human body in terms of measuring the present current. One-dimensional electrical direct current in a thin, straight conductor could also be applied to the visualization of instruments in interventional MRI procedures. Currents below 100 mA would be sufficient to create distinct artifacts, at least under simplified conditions (homogeneous background etc.).
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Próteses e Implantes
/
Processamento de Imagem Assistida por Computador
/
Imageamento por Ressonância Magnética
/
Terapia por Estimulação Elétrica
/
Artefatos
/
Cobre
/
Condutividade Elétrica
Tipo de estudo:
Diagnostic_studies
Limite:
Humans
Idioma:
En
Revista:
Z Med Phys
Assunto da revista:
RADIOTERAPIA
Ano de publicação:
2011
Tipo de documento:
Article
País de afiliação:
Alemanha
País de publicação:
Alemanha