Changing from analog to digital images: does it affect the accuracy of alignment measurements of the lower extremity?
Lohman, Martina; Tallroth, Kaj; Kettunen, Jyrki A; Remes, Ville.
Afiliação
  • Lohman M; Helsinki University Central Hospital, Helsinki, Finland. martina.lohman@hus.fi
Acta Orthop ; 82(3): 351-5, 2011 Jun.
Article em En | MEDLINE | ID: mdl-21619504
ABSTRACT
BACKGROUND AND

PURPOSE:

Medical imaging has changed from analog films to digital media. We examined and compared the accuracy of orthopedic measurements using different media.

METHODS:

Before knee arthroplasty, full-length standing radiographs of 52 legs were obtained. The mechanical axis (MA), tibio-femoral angle (TFA), and femur angle (FA) were measured and analyzed twice, by 2 radiologists, using (1) true-size films, (2) short films, (3) a digital high-resolution workstation, and (4) a web-based personal computer. The agreement between the 4 media was evaluated using the Bland-Altman method (limits of agreement) using the true-size films as a reference standard.

RESULTS:

The mean differences in measurements between the traditional true-size films and the 3 other methods were small for MA -0.20 to 0.07 degrees, and for TFA -0.02 to 0.18 degrees. Also, the limits of agreement between the traditional true-size films and the three other methods were small.

INTERPRETATION:

The agreement of the alignment measurements across the 4 different media was good. Orthopedic angles can be measured as accurately from analog films as from digital screens, regardless of film or monitor size.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Mau Alinhamento Ósseo / Artroplastia do Joelho / Extremidade Inferior Tipo de estudo: Diagnostic_studies / Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Acta Orthop Assunto da revista: ORTOPEDIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Finlândia

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Mau Alinhamento Ósseo / Artroplastia do Joelho / Extremidade Inferior Tipo de estudo: Diagnostic_studies / Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Acta Orthop Assunto da revista: ORTOPEDIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Finlândia