Feasibility of percutaneous implantation of transapical endocardial left ventricular pacing electrode for cardiac resynchronization therapy.
Kassai, Imre; Friedrich, Orsolya; Ratnatunga, Chandi; Betts, Timothy R; Mihálcz, Attila; Szili-Török, Tamás.
Afiliação
  • Kassai I; Department of Cardiac Surgery, Gottsegen Hungarian Institute of Cardiology, Budapest, Hungary. imrekassai@hotmail.com
Europace ; 13(11): 1653-7, 2011 Nov.
Article em En | MEDLINE | ID: mdl-21712260
ABSTRACT
Failure of coronary sinus lead implantation for resynchronization therapy requires alternative approaches. For such events we have developed a transapical implantation technique as a feasible alternative. We report the outcome of this technique and its evolution from a minithoracotomy to a percutaneous approach. Twenty patients underwent alternative resynchronization therapy with transapical endocardial left ventricular (LV) pacing lead implantation in a multicentre, international study between October 2007 and March 2010. Eighteen patients underwent minithoracotomy and transapical puncture under direct observation. Two recent patients had transthoracic echocardiography-guided percutaneous apical puncture to enter the LV cavity. A 19 or 21 ga needle and two-stage Seldinger dilatation with 4 and 7 Fr sheaths were then used to introduce the lead. In the two patients with closed-chest insertion of the electrode there was no puncture related bleeding or lung damage. Lead dislocation occurred in two minithoracotomy patients. Repositioning was performed without re-opening the pleural cavity. One patient developed right-sided implanted cardiac defibrillator lead endocarditis requiring complete system removal. Twelve patients have >1 year follow-up; all have sustained and significant improvement in LV dimensions (diastolic Δ4.2 ± 2.9, systolic Δ7.2 ± 5.8 mm), ejection fraction (Δ9.5 ± 9.6%), and functional status (Δ1.1 ± 0.3). Transapical placement of LV endocardial pacing lead is an effective alternative strategy for cardiac resynchronization. A closed-chest, percutaneous approach is feasible and should offer even less invasive intervention.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Disfunção Ventricular Esquerda / Eletrodos Implantados / Endocárdio / Terapia de Ressincronização Cardíaca / Dispositivos de Terapia de Ressincronização Cardíaca / Insuficiência Cardíaca / Ventrículos do Coração Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Revista: Europace Assunto da revista: CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Hungria

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Disfunção Ventricular Esquerda / Eletrodos Implantados / Endocárdio / Terapia de Ressincronização Cardíaca / Dispositivos de Terapia de Ressincronização Cardíaca / Insuficiência Cardíaca / Ventrículos do Coração Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Revista: Europace Assunto da revista: CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Hungria