Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6')-Ib-cr.
Calvo, Jorge; Cano, María Eliecer; Pitart, Cristina; Marco, Francesc; Rodríguez-Martínez, José Manuel; Pascual, Alvaro; Martínez-Martínez, Luis.
Afiliação
  • Calvo J; Service of Microbiology, University Hospital Marqués de Valdecilla, Av. Valdecilla s/n, 39008 Santander, Spain. jcalvo@humv.es.
J Clin Microbiol ; 49(9): 3343-5, 2011 Sep.
Article em En | MEDLINE | ID: mdl-21775549
ABSTRACT
The accuracy of the MicroScan WalkAway, BD Phoenix, and Vitek-2 systems for susceptibility testing of quinolones and aminoglycosides against 68 enterobacteria containing qnrB, qnrS, and/or aac(6 ')-Ib-cr was evaluated using reference microdilution. Overall, one very major error (0.09%), 6 major errors (0.52%), and 45 minor errors (3.89%) were noted.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Automação / Enterobacteriaceae / Antibacterianos Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Revista: J Clin Microbiol Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Espanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Automação / Enterobacteriaceae / Antibacterianos Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Revista: J Clin Microbiol Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Espanha