Fast heating induced impulse halogenation of refractory sample components in electrothermal atomic absorption spectrometry by direct injection of a liquid halogenating agent.
Talanta
; 85(3): 1253-9, 2011 Sep 15.
Article
em En
| MEDLINE
| ID: mdl-21807179
A novel electrothermal atomic absorption spectrometry (ETAAS) method was developed for the halogenation of refractory sample components (Er, Nd and Nb) of lithium niobate (LiNbO(3)) and bismuth tellurite (Bi(2)TeO(5)) optical single crystals to overcome memory effects and carry-over. For this purpose, the cleaning step of a regular graphite furnace heating program was replaced with a halogenation cycle. In this cycle, after the graphite tube cooled to room temperature, a 20 µL aliquot of liquid carbon tetrachloride (CCl(4)) was dispensed with a conventional autosampler into the graphite tube. The CCl(4) was partially dried at 80°C under the mini-flow (40 cm(3) min(-1)) condition of the Ar internal furnace gas (IFG), then the residue was decomposed (pyrolyzed) by fast furnace heating at 1900-2100°C under interrupted flow of the IFG. This step was followed by a clean-out stage at 2100°C under the maximum flow of the IFG. The advantage of the present method is that it does not require any alteration to the graphite furnace gas supply system in contrast to most of the formerly introduced halogenation techniques. The effectiveness of the halogenation method was verified with the determination of Er and Nd dopants in the optical crystals. In these analyses, a sensitivity decrease was observed, which was likely due to the enhanced deterioration of the graphite tube surface. Therefore, the application of mathematical correction (resloping) of the calibration was also required. The calibration curves were linear up to 1.5 and 10 µmol L(-1) for Er and Nd, respectively. Characteristic masses of 18 and 241 pg and the limit of detection (LOD) values of 0.017 and 0.27 µmol L(-1) were found for Er and Nd, respectively. These LOD data correspond to 0.68 µmol mol(-1) Er and 11 µmol mol(-1) Nd in solid bismuth tellurite samples. The analytical results were compared with those obtained by a conventional ETAAS method and validated with X-ray fluorescence spectrometry analysis.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Espectrofotometria Atômica
/
Tetracloreto de Carbono
/
Halogenação
/
Calefação
Idioma:
En
Revista:
Talanta
Ano de publicação:
2011
Tipo de documento:
Article
País de afiliação:
Hungria
País de publicação:
Holanda