An efficient and inexpensive refrigerated LC system for H/D exchange mass spectrometry.
Keppel, Theodore R; Jacques, Martin E; Young, Robert W; Ratzlaff, Kenneth L; Weis, David D.
Afiliação
  • Keppel TR; Department of Chemistry, University of Kansas, Lawrence, KS 66045, USA.
J Am Soc Mass Spectrom ; 22(8): 1472-6, 2011 Aug.
Article em En | MEDLINE | ID: mdl-21953202
Loss of deuterium label during the LC step in amide hydrogen/deuterium exchange mass spectrometry (H/D-MS) is minimized by maintaining an acidic mobile phase pH and low temperature (pH 2.5, 0 °C). Here we detail the construction and performance of a low-cost, thermoelectrically refrigerated enclosure to house high-performance liquid chromatography (HPLC) components and cool mobile phases. Small volume heat exchangers rapidly decrease mobile phase temperature and keep the temperature stable to ±0.2 °C. Using a superficially porous reversed-phase column, we obtained excellent chromatographic performance in the separation of peptides with a median peak width of 4.4 s. Average deuterium recovery was 80.2% with an average relative precision of 0.91%.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Refrigeração / Espectrometria de Massas / Medição da Troca de Deutério / Cromatografia de Fase Reversa Idioma: En Revista: J Am Soc Mass Spectrom Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Refrigeração / Espectrometria de Massas / Medição da Troca de Deutério / Cromatografia de Fase Reversa Idioma: En Revista: J Am Soc Mass Spectrom Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos