An efficient and inexpensive refrigerated LC system for H/D exchange mass spectrometry.
J Am Soc Mass Spectrom
; 22(8): 1472-6, 2011 Aug.
Article
em En
| MEDLINE
| ID: mdl-21953202
Loss of deuterium label during the LC step in amide hydrogen/deuterium exchange mass spectrometry (H/D-MS) is minimized by maintaining an acidic mobile phase pH and low temperature (pH 2.5, 0 °C). Here we detail the construction and performance of a low-cost, thermoelectrically refrigerated enclosure to house high-performance liquid chromatography (HPLC) components and cool mobile phases. Small volume heat exchangers rapidly decrease mobile phase temperature and keep the temperature stable to ±0.2 °C. Using a superficially porous reversed-phase column, we obtained excellent chromatographic performance in the separation of peptides with a median peak width of 4.4 s. Average deuterium recovery was 80.2% with an average relative precision of 0.91%.