Prospective comparison of discrimination algorithms to prevent inappropriate ICD therapy: primary results of the Rhythm ID Going Head to Head Trial.
Gold, Michael R; Ahmad, Saleem; Browne, Kevin; Berg, Kellie Chase; Thackeray, Lisa; Berger, Ronald D.
Afiliação
  • Gold MR; Medical University of South Carolina, Charleston, South Carolina 29425, USA. goldmr@musc.edu
Heart Rhythm ; 9(3): 370-7, 2012 Mar.
Article em En | MEDLINE | ID: mdl-21978966
ABSTRACT

BACKGROUND:

Inappropriate therapy for supraventricular arrhythmias remains a significant source of morbidity in implantable cardioverter-defibrillator (ICD) recipients.

OBJECTIVE:

The Rhythm ID Goes Head to Head Trial (RIGHT) was designed to compare rhythm discrimination and inappropriate therapies among patients with ICDs from 2 manufacturers.

METHODS:

Patients with standard ICD indications were randomized to receive a Guidant VITALITY 2 with Rhythm ID or selective Medtronic pulse generators using the Enhanced PR Logic or Wavelet discrimination algorithms. A single- or dual-chamber device was implanted based on clinical indications and programmed in 2 detection zones with detection enhancements enabled for rates between 150 and 200 bpm. Algorithm performance was compared between randomization groups, stratified by single or dual chamber, for the primary end point of first inappropriate therapy (shock or antitachycardia pacing) for supraventricular arrhythmias.

RESULTS:

There were 1962 patients enrolled and followed for 18.3 ± 9.2 months, with no difference in all-cause mortality between groups. There were 3973 treated episodes where electrograms were available and adjudicated. The primary end point of inappropriate therapy occurred in 246 of 985 VITALITY 2 patients vs 187 of 977 specific Medtronic ICD patients (hazard ratio = 1.34; confidence interval = 1.11-1.62; P = .003). Differences in inappropriate therapy were confined to single-chamber ICDs. Inappropriate shocks were more frequent in VITALITY 2 ICDs (hazard ratio = 1.63; confidence interval = 1.29-2.06; P < .001), with most therapies and performance differences occurring at slower rhythms (rates < 175 bpm).

CONCLUSION:

Rhythm discrimination performed better in the specific Medtronic than in VITALITY 2 ICDs evaluated, particularly for single-chamber devices. Inappropriate therapies, and differences in performance, may be reduced with the use of rate cutoff above 175 bpm.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Taquicardia Supraventricular / Cardioversão Elétrica / Estimulação Cardíaca Artificial / Taquicardia Ventricular / Desfibriladores Implantáveis Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies / Prognostic_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Heart Rhythm Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Taquicardia Supraventricular / Cardioversão Elétrica / Estimulação Cardíaca Artificial / Taquicardia Ventricular / Desfibriladores Implantáveis Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies / Prognostic_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Heart Rhythm Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Estados Unidos