Accuracy and precision of vessel area assessment: manual versus automatic lumen delineation based on full-width at half-maximum.
Merkx, Maarten A G; Bescós, Javier Oliván; Geerts, Liesbeth; Bosboom, E Mariëlle H; van de Vosse, Frans N; Breeuwer, Marcel.
Afiliação
  • Merkx MA; Maastricht University Medical Center, Department of Biomedical Engineering, Maastricht, The Netherlands. m.a.g.merkx@alumnus.tue.nl
J Magn Reson Imaging ; 36(5): 1186-93, 2012 Nov.
Article em En | MEDLINE | ID: mdl-22826150
ABSTRACT

PURPOSE:

To evaluate the accuracy and precision of manual and automatic blood vessel diameter measurements, a quantitative comparison was conducted, using both phantom and clinical 3D magnetic resonance angiography (MRA) data. Since diameters are often manually measured, which likely is influenced by operator dependency, automatic lumen delineation, based on the full-width at half-maximum (FWHM), could improve these measurements. MATERIALS AND

METHODS:

Manual and automatic diameter assessments were compared, using MRA data from a vascular phantom (geometry obtained with µCT) and clinical MRA data. The diameters were manually assessed by 15 MRA experts, using both caliper and contour tools. To translate the experimental results to clinical practice, the precision obtained using phantom data was compared to the precision obtained with clinical data.

RESULTS:

A diameter error <10% was obtained with resolutions above 2, 3, and 5 pixels/diameter for the automatic FWHM, contour, and caliper methods, respectively. Using phantom data, precision of the manual methods was low (error >20%), even at high resolutions, while precision for the automatic method was high (error <3%) when using more than 2 pixels/diameter. A similar trend was found with clinical data.

CONCLUSION:

The results obtained clearly demonstrate improvement in the accuracy and precision of vessel diameter measurements with use of the automatic FWHM-based method.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artérias / Algoritmos / Reconhecimento Automatizado de Padrão / Interpretação de Imagem Assistida por Computador / Angiografia por Ressonância Magnética / Imageamento Tridimensional / Doença Arterial Periférica Tipo de estudo: Diagnostic_studies / Guideline Limite: Humans Idioma: En Revista: J Magn Reson Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artérias / Algoritmos / Reconhecimento Automatizado de Padrão / Interpretação de Imagem Assistida por Computador / Angiografia por Ressonância Magnética / Imageamento Tridimensional / Doença Arterial Periférica Tipo de estudo: Diagnostic_studies / Guideline Limite: Humans Idioma: En Revista: J Magn Reson Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Holanda