Prevention of a wrong-location misadministration through the use of an intradepartmental incident learning system.
Ford, Eric C; Smith, Koren; Harris, Kendra; Terezakis, Stephanie.
Afiliação
  • Ford EC; Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University, Baltimore, MD 21231, USA.
Med Phys ; 39(11): 6968-71, 2012 Nov.
Article em En | MEDLINE | ID: mdl-23127090
ABSTRACT

PURPOSE:

A series of examples are presented in which potential errors in the delivery of radiation therapy were prevented through use of incident learning. These examples underscore the value of reporting near miss incidents.

METHODS:

Using a departmental incident learning system, eight incidents were noted over a two-year period in which fields were treated "out-of-sequence," that is, fields from a boost phase were treated, while the patient was still in the initial phase of treatment. As a result, an error-prevention policy was instituted in which radiation treatment fields are "hidden" within the oncology information system (OIS) when they are not in current use. In this way, fields are only available to be treated in the intended sequence and, importantly, old fields cannot be activated at the linear accelerator control console.

RESULTS:

No out-of-sequence treatments have been reported in more than two years since the policy change. Furthermore, at least three near-miss incidents were detected and corrected as a result of the policy change. In the first two, the policy operated as intended to directly prevent an error in field scheduling. In the third near-miss, the policy operated "off target" to prevent a type of error scenario that it was not directly intended to prevent. In this incident, an incorrect digitally reconstructed radiograph (DRR) was scheduled in the OIS for a patient receiving lung cancer treatment. The incorrect DRR had an isocenter which was misplaced by approximately two centimeters. The error was a result of a field from an old plan being scheduled instead of the intended new plan. As a result of the policy described above, the DRR field could not be activated for treatment however and the error was discovered and corrected. Other quality control barriers in place would have been unlikely to have detected this error.

CONCLUSIONS:

In these examples, a policy was adopted based on incident learning, which prevented several errors, at least one of which was potentially severe. These examples underscore the need for a rigorous, systematic incident learning process within each clinic. The experiences reported in this technical note demonstrate the value of near-miss incident reporting to improve patient safety.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bases de Dados Factuais / Erros de Configuração em Radioterapia / Departamentos Hospitalares Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bases de Dados Factuais / Erros de Configuração em Radioterapia / Departamentos Hospitalares Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Estados Unidos