Reintervention for distal stent graft-induced new entry after endovascular repair with a stainless steel-based device in aortic dissection.
Weng, Shih-Hsien; Weng, Chi-Feng; Chen, Wei-Yuan; Huang, Chun-Yang; Chen, I-Ming; Chen, Chun-Ku; Hsu, Chiao-Po; Shih, Chun-Che.
Afiliação
  • Weng SH; Division of Cardiovascular Surgery, Department of Surgery, Taipei Veterans General Hospital, Institutes of Clinical Medicine, National Yang-Ming University School of Medicine, Taipei, Taiwan.
J Vasc Surg ; 57(1): 64-71, 2013 Jan.
Article em En | MEDLINE | ID: mdl-23141675
ABSTRACT

OBJECTIVE:

Stent graft-induced new entry (SINE) has been increasingly observed after thoracic endovascular aortic repair (TEVAR) for aortic dissection. We investigated the mechanism of late distal SINE, prevention strategies, proper size selection of the stent graft, and implantation sequence.

METHODS:

From November 2006 to May 2011, 99 patients with aortic dissection underwent TEVAR with Zenith TX2 stent grafts (Cook, Bloomington, Ind) at our center. Among them, 27 distal SINEs were recognized. Eight of these patients with complicated distal SINE required intervention with new distal endografts, and all were enrolled for further analysis.

RESULTS:

Eight of the 27 patients with distal SINE underwent a secondary endograft procedure from February 2011 to July 2011. All were successfully treated without any complications or deaths. A high taper ratio (35%±11%) and excessive oversizing of the true lumen area at the distal stent level (293%±76%) were noted among these patients.

CONCLUSIONS:

The incidence of distal SINE seemed to be high; however, there were also low rates of death and complications after TEVAR for aortic dissection using stainless steel-based stent grafts. Complicated distal SINE can successfully be resolved by distal endograft implantation. Excessive oversizing of the distal stent graft, as measured by the true lumen area, may be a significant factor causing delayed distal SINE. Precise size selection is crucial for the distal end of the stent, especially for high taper ratio dissection pathology in which the implantation sequence of a distal small-sized stent graft first might be considered to prevent future distal SINE.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Complicações Pós-Operatórias / Aço Inoxidável / Prótese Vascular / Stents / Aneurisma da Aorta Torácica / Implante de Prótese Vascular / Procedimentos Endovasculares / Dissecção Aórtica Tipo de estudo: Etiology_studies / Incidence_studies / Prognostic_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Vasc Surg Assunto da revista: ANGIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Taiwan

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Complicações Pós-Operatórias / Aço Inoxidável / Prótese Vascular / Stents / Aneurisma da Aorta Torácica / Implante de Prótese Vascular / Procedimentos Endovasculares / Dissecção Aórtica Tipo de estudo: Etiology_studies / Incidence_studies / Prognostic_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Vasc Surg Assunto da revista: ANGIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Taiwan