Source effects in analyzer-based X-ray phase contrast imaging with conventional sources.
Hönnicke, M G; Manica, J; Mazzaro, I; Cusatis, C; Huang, X-R.
Afiliação
  • Hönnicke MG; Universidade Federal da Integração Latino-Americana, 85867-970 Foz do Iguaçu, PR, Brazil. marcelo.honnicke@unila.edu.br
Rev Sci Instrum ; 83(11): 113702, 2012 Nov.
Article em En | MEDLINE | ID: mdl-23206066
ABSTRACT
Several recent papers have shown the implementation of analyzer based X-ray phase contrast imaging (ABI) with conventional X-ray sources. The high flux is always a requirement to make the technique useful for bio-medical applications. Here, we present and discuss three important parameters, which need to be taken into account, when searching for the high flux ABI anisotropic magnification, double image, and source size spread due to intrinsic dispersive diffraction by asymmetrically cut crystals. These parameters, if not well optimized, may cause important features in the acquired images which can mislead the interpretation. A few ways to minimize these effects are implemented and discussed, including some experimental results.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Brasil