Novel electrode configuration for highly linear impedance pneumography.
Seppä, Ville-Pekka; Hyttinen, Jari; Uitto, Marko; Chrapek, Wojciech; Viik, Jari.
Afiliação
  • Seppä VP; Department of Electronics and Communications Engineering, Tampere University of Technology, Tampere, Finland. ville-pekka.seppa@tut.fi
Biomed Tech (Berl) ; 58(1): 35-8, 2013 Feb.
Article em En | MEDLINE | ID: mdl-23348215
ABSTRACT
Impedance pneumography (IP) is a non-invasive respiration measurement technique. Emerging applications of IP in respiratory medicine use the measured signal to monitor pulmonary flow and volume parameters related to airway obstruction during tidal breathing (TB). This requires a high impedance change (ΔZ)-to-lung volume change (ΔV) linearity. Four potential electrode configurations were tested on 10 healthy subjects. Only the novel configuration where the electrodes were placed in both the thorax and the arms yielded a highly linear ΔZ/ΔV in all subjects. The presented electrode configuration may expand the clinical use of IP from the conventional tidal volume estimation to flow measurement.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espirometria / Cardiografia de Impedância / Volume de Ventilação Pulmonar / Eletrodos Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Biomed Tech (Berl) Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Finlândia

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espirometria / Cardiografia de Impedância / Volume de Ventilação Pulmonar / Eletrodos Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Biomed Tech (Berl) Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Finlândia