Polyurethane does not protect better than polyvinyl cuffed tracheal tubes from microaspirations.
Bulpa, P; Evrard, P; Bouhon, S; Schryvers, F; Jamart, J; Michaux, I; Dive, A; Vander Borght, T; Krug, B.
Afiliação
  • Bulpa P; Intensive Care Unit, CHU Mont-Godinne, Université Catholique de Louvain, 5530 Yvoir, Belgium. pierre.bulpa@uclouvain.be
Minerva Anestesiol ; 79(5): 498-503, 2013 May.
Article em En | MEDLINE | ID: mdl-23511349
BACKGROUND: Mechanically ventilated patients are prone to develop ventilator associated pneumonia due to microaspirations of subglottic secretions around the endotracheal tube cuff (usually constructed of polyvinyl material). A novel polyurethane cuff has been designed to minimize these leakages. The aim of the study was to compare the tracheal sealing capacities between the two tubes. METHODS: Twenty-nine consecutive patients from whom tracheal intubation was necessary as part of their care were randomized to receive either a polyvinyl HI-LO Evac® or a polyurethane SEALGUARD Evac® endotracheal tube. Patients requiring emergency intubation, with unstable hemodynamics or history of tracheal/laryngeal disease were excluded. For the entire study, cuff pressure was set at 30 cmH2O, and ventilator parameters were adjusted for a plateau pressure ≤30 cmH2O; Patients were fasting, placed in a strict 45° position during 12 hours and sedated if needed. After injection of 74 MBq 99mTc-DTPA diluted in 5 mL 0.9% NaCl just above the cuff, tracheal radioactivity was assessed sequentially (hourly from T0 to T6, then T8 and T12 hours) using a scintillation camera. RESULTS: Sixteen polyurethane and 13 polyvinyl tubes were compared. Leakages were observed in 11/29 patients (38%) (5/16 polyurethane and 6/13 polyvinyl tubes [P=NS]). Leakages occurred more often in female (7/8) than in male patients (4/21) (P<0.001). Microaspirations were decreased with larger tubes (size 9 vs. ≤8.5: 24% vs. 75%; P=0.01), whatever the cuff membrane. CONCLUSION: These preliminary results suggest that both tubes are poorly effective in preventing microaspirations.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pneumonia Aspirativa / Poliuretanos / Polivinil / Intubação Intratraqueal / Manufaturas Tipo de estudo: Clinical_trials / Observational_studies / Risk_factors_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Minerva Anestesiol Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Bélgica País de publicação: Itália
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pneumonia Aspirativa / Poliuretanos / Polivinil / Intubação Intratraqueal / Manufaturas Tipo de estudo: Clinical_trials / Observational_studies / Risk_factors_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Minerva Anestesiol Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Bélgica País de publicação: Itália