MR guided applicator reconstruction for brachytherapy of cervical cancer using the novel titanium Rotterdam applicator.
Radiother Oncol
; 107(1): 88-92, 2013 Apr.
Article
em En
| MEDLINE
| ID: mdl-23639579
ABSTRACT
A novel model of the titanium Rotterdam tandem and ovoid applicator is presented. As titanium produces artefacts in MR images, an MR sequence was sought and optimised for visualisation and accurate applicator reconstruction. The mean inter-observer (8 observers) variability for four patients was only 0.7 mm (maximum 1.7 mm).
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Braquiterapia
/
Neoplasias do Colo do Útero
/
Imagem por Ressonância Magnética Intervencionista
/
Radioterapia Guiada por Imagem
Limite:
Female
/
Humans
Idioma:
En
Revista:
Radiother Oncol
Ano de publicação:
2013
Tipo de documento:
Article
País de afiliação:
Holanda