Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial.
Gasparini, Maurizio; Proclemer, Alessandro; Klersy, Catherine; Kloppe, Axel; Lunati, Maurizio; Ferrer, José Bautista Martìnez; Hersi, Ahmad; Gulaj, Marcin; Wijfels, Maurits C E F; Santi, Elisabetta; Manotta, Laura; Arenal, Angel.
Afiliação
  • Gasparini M; Electrophysiology and Pacing Unit, Humanitas Clinical and Research Center, Via Manzoni 56, 20089 Rozzano (MI), Italy. maurizio.gasparini@humanitas.it
JAMA ; 309(18): 1903-11, 2013 May 08.
Article em En | MEDLINE | ID: mdl-23652522
IMPORTANCE: Using more intervals to detect ventricular tachyarrhythmias has been associated with reducing unnecessary implantable cardioverter-defibrillator (ICD) therapies. OBJECTIVE: To determine whether using 30 of 40 intervals to detect ventricular arrhythmias (VT) (long detection) during spontaneous fast VT episodes reduces antitachycardia pacing (ATP) and shock delivery more than 18 of 24 intervals (standard detection). DESIGN, SETTING, AND PARTICIPANTS: Randomized, single-blind, parallel-group trial that enrolled 1902 primary and secondary prevention patients (mean [SD] age, 65 [11] years; 84% men; 75% primary prevention ICD) with ischemic and nonischemic etiology undergoing first ICD implant at 1 of 94 international centers (March 2008-December 2010). INTERVENTIONS: Patients were randomized 1:1 to programming with long- (n = 948) or standard-detection (n = 954) intervals. MAIN OUTCOMES AND MEASURES: Total number of ATPs and shocks delivered for all episodes (primary outcomes) and inappropriate shocks, mortality, and syncopal rate (secondary outcomes). RESULTS: During a median follow-up of 12 months (interquartile range, 11-13), long-detection group had 346 delivered therapies (42 therapies per 100 person-years, 95% CI, 38-47) vs 557 in the standard-detection group (67 therapies per 100 person-years [95% CI, 62-73]; incident rate ratio [IRR], 0.63 [95% CI, 0.51-0.78]; P < .001). The long- vs the standard-detection group experienced 23 ATPs per 100 person-years (95% CI, 20-27) vs 37 ATPs per 100 person-years (95% CI, 33-41; IRR, 0.58 [95% CI, 0.47-0.72]; P < .001); 19 shocks per 100 person-years (95% CI, 16-22) vs 30 shocks per 100 person-years (95% CI, 26-34; IRR, 0.77 [95% CI, 0.59-1.01]; P = .06), with a significant difference in the probability of therapy occurrence (P < .001); and a reduction in first occurrence of inappropriate shock (5.1 per 100 patient-years [95% CI, 3.7-6.9] vs 11.6 [95% CI, 9.4-14.1]; IRR, 0.55 [95% CI, 0.36-0.85]; P = .008). Mortality (5.5 [95% CI, 4.0-7.2] vs 6.3 [95% CI, 4.8-8.2] per 100 patient-years; HR, 0.87; P = .50) and arrhythmic syncope rates (3.1 [95% CI, 2.6-4.6] vs 1.9 [95% CI, 1.1-3.1] per 100 patient-years; IRR, 1.60 [95% CI, 0.76-3.41]; P = .22) did not differ significantly between groups. CONCLUSIONS AND RELEVANCE: Among patients receiving an ICD, the use of a long- vs standard-detection interval resulted in a lower rate of ATP and shocks, and inappropriate shocks. This programming strategy may be an appropriate alternative. TRIAL REGISTRATION: clinicaltrials.gov Identifier: NCT00617175.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Cardioversão Elétrica / Desfibriladores Implantáveis Tipo de estudo: Clinical_trials / Diagnostic_studies / Prognostic_studies Aspecto: Patient_preference Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: JAMA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Itália País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Cardioversão Elétrica / Desfibriladores Implantáveis Tipo de estudo: Clinical_trials / Diagnostic_studies / Prognostic_studies Aspecto: Patient_preference Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: JAMA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Itália País de publicação: Estados Unidos