Your browser doesn't support javascript.
loading
Long-term survival of the uncemented Balgrist total hip replacement cup.
Gallo, Jiri; Lostak, Jiri; Langova, Katerina.
Afiliação
  • Gallo J; Department of Orthopaedics, Faculty of Medicine and Dentistry, Palacky University Olomouc, University Hospital, I. P. Pavlova 6, 77520, Olomouc, Czech Republic. jiri.gallo@volny.cz
Int Orthop ; 37(8): 1449-56, 2013 Aug.
Article em En | MEDLINE | ID: mdl-23744502
ABSTRACT

PURPOSE:

The purpose of this study was to evaluate the long-term clinical and radiographic outcomes of the Balgrist total hip replacement cup.

METHODS:

We report the results of a retrospective review of 195 total hip prostheses with Balgrist cup implanted in 178 patients. There were 59 men and 119 women with a mean age at surgery of 52.3 years (range, 19-72). Cox regression was used to determine the influence of independent variables on the survivorship of the prosthesis.

RESULTS:

There were 117 (66 %) of 178 patients, and 131 (67 %) of 195 hips available for assessment at a mean follow-up of 17.4 years. A total of 21 patients (22 hips) were lost from follow-up (12 %). A total of 25 patients (27 hips) required some type of re-operation during the follow-up period. Eight of 27 re-operations (30 %) were directly caused by the Balgrist cup failure due to fatigue fracture of the metallic shell (n = 5; 19 %), polyethylene wear (n = 2; 7 %) and aseptic loosening (n = 1; 4 %). The mean polyethylene wear rate was 0.068 mm/year (range, 0.008-0.230; SD = 0.043). The risk of re-operation for any reason was decreased with dysplastic hip as a primary diagnosis (hazard ratio; HR = 0.250; 95 % CI 0.086-0.725) and with greater age at the time of surgery (HR = 0.923; 95 % CI 0.880-0.969). The 19-year survivorship with re-operation for any reason was 75.6 % (95 % CI 67.0-84.2). The 19-year survivorship with re-operation for the Balgrist cup failure was 90.5 % (95 % CI 83.4-97.6).

CONCLUSIONS:

The Balgrist cup continues to provide excellent clinical and radiological outcomes. This is associated at least in part with a low polyethylene wear rate. The main reason for the Balgrist cup failure is fatigue fracture of the metallic shell.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Osteoartrite do Quadril / Artroplastia de Quadril / Luxação do Quadril / Articulação do Quadril / Prótese de Quadril Tipo de estudo: Diagnostic_studies / Etiology_studies / Evaluation_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Int Orthop Ano de publicação: 2013 Tipo de documento: Article País de afiliação: República Tcheca

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Osteoartrite do Quadril / Artroplastia de Quadril / Luxação do Quadril / Articulação do Quadril / Prótese de Quadril Tipo de estudo: Diagnostic_studies / Etiology_studies / Evaluation_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Int Orthop Ano de publicação: 2013 Tipo de documento: Article País de afiliação: República Tcheca