[Comparison of dose calculation algorithms in stereotactic radiation therapy in lung].
Tomiyama, Yuki; Araki, Fujio; Kanetake, Nagisa; Shimohigashi, Yoshinobu; Tominaga, Hirofumi; Sakata, Jyunichi; Oono, Takeshi; Kouno, Tomohiro; Hioki, Kazunari.
Afiliação
  • Tomiyama Y; Graduate School of Health Science, Kumamoto University.
Nihon Hoshasen Gijutsu Gakkai Zasshi ; 69(6): 663-8, 2013 Jun.
Article em Ja | MEDLINE | ID: mdl-23782779
Dose calculation algorithms in radiation treatment planning systems (RTPSs) play a crucial role in stereotactic body radiation therapy (SBRT) in the lung with heterogeneous media. This study investigated the performance and accuracy of dose calculation for three algorithms: analytical anisotropic algorithm (AAA), pencil beam convolution (PBC) and Acuros XB (AXB) in Eclipse (Varian Medical Systems), by comparison against the Voxel Monte Carlo algorithm (VMC) in iPlan (BrainLab). The dose calculations were performed with clinical lung treatments under identical planning conditions, and the dose distributions and the dose volume histogram (DVH) were compared among algorithms. AAA underestimated the dose in the planning target volume (PTV) compared to VMC and AXB in most clinical plans. In contrast, PBC overestimated the PTV dose. AXB tended to slightly overestimate the PTV dose compared to VMC but the discrepancy was within 3%. The discrepancy in the PTV dose between VMC and AXB appears to be due to differences in physical material assignments, material voxelization methods, and an energy cut-off for electron interactions. The dose distributions in lung treatments varied significantly according to the calculation accuracy of the algorithms. VMC and AXB are better algorithms than AAA for SBRT.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Radiocirurgia / Pulmão Tipo de estudo: Health_economic_evaluation Limite: Humans Idioma: Ja Revista: Nihon Hoshasen Gijutsu Gakkai Zasshi Ano de publicação: 2013 Tipo de documento: Article País de publicação: Japão
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Radiocirurgia / Pulmão Tipo de estudo: Health_economic_evaluation Limite: Humans Idioma: Ja Revista: Nihon Hoshasen Gijutsu Gakkai Zasshi Ano de publicação: 2013 Tipo de documento: Article País de publicação: Japão