Initial and pioneer experience of transcatheter aortic valve implantation (Inovare) through femoral or iliac artery.
Pontes, José Carlos Dorsa Vieira; Duarte, João Jackson; Silva, Augusto Daige da; Gardenal, Neimar; Dias, Amaury Mont'Serrat Ávila Souza; Benfatti, Ricardo Adala; Silva, Guilherme Viotto Rodrigues da; Benfatti, Amanda Ferreira Carli.
Afiliação
  • Pontes JC; University Hospital of Federal University of Mato Grosso do Sul (UFMS), Campo Grande, MS, Brazil. carlosdorsa@uol.com.br
Rev Bras Cir Cardiovasc ; 28(2): 208-16, 2013 Jun.
Article em En, Pt | MEDLINE | ID: mdl-23939317
ABSTRACT

OBJECTIVE:

This paper demonstrates the initial and pioneering experience implant of the Inovare prosthesis implant through transfemoral or iliac artery route.

METHODS:

Six patients underwent transcatheter aortic valve implantation. The access was femoral or iliac through which the delivery device, a latex balloon catheter with the crimped prosthesis, was inserted. Through the femoral introducer 24 Fr Gore® DrySeal sheath, an extra stiff guide wire with non-traumatic tip was positioned in the left ventricle by passing through the valve ring. After balloon valvuloplasty, in cases of native valve stenosis, the prosthesis implantation was performed after hypotension induced by tachycardia and controlled by temporary pacemaker. The valve positioning was guided by TEE (transesophageal ecocardiography) and fluoroscopy, aiming to position a third of the length of the prosthesis into the left ventricle cavity.

RESULTS:

The successful valve implantation was possible in six cases. There was no need of conversion to open surgery due to inability to access or graft migration. There were no intraoperative or hospital deaths. We observed a significant reduction in the mean gradient of 66.84115.46 mmHg to 19.74110.61 mm Hg postoperatively (P=0.002), a reduction of 70.46%.

CONCLUSION:

Inovare prosthesis, implanted by femoral or iliac artery was feasible, and determined adequate hemodynamic performance in the postoperative follow-up, showing no mortality in this small series.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Cateterismo Cardíaco / Implante de Prótese de Valva Cardíaca / Artéria Femoral / Artéria Ilíaca Tipo de estudo: Etiology_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En / Pt Revista: Rev Bras Cir Cardiovasc Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Brasil

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Cateterismo Cardíaco / Implante de Prótese de Valva Cardíaca / Artéria Femoral / Artéria Ilíaca Tipo de estudo: Etiology_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En / Pt Revista: Rev Bras Cir Cardiovasc Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Brasil