Construction and in vitro test of a new electrode for dentin resistance measurement.
Biomed Tech (Berl)
; 58(5): 469-74, 2013 Oct.
Article
em En
| MEDLINE
| ID: mdl-24021589
It is necessary to reduce the tooth substance before treating a tooth with a dental crown. The preparation often requires reduction of the dentin. This results in a dentin wound and a thinner substance over the pulp, increases the risk of inflammation, and could result in necrosis of the pulp. To give the dentist information about the amount of dentin over the pulp during preparation, the Prepometer was developed. The function of this device is based on the measurement of the electric resistance of the tooth substance. The measuring behavior of the first-generation Prepometer is characterized by smaller values of electric resistance before reaching full contact of the measuring head to the dentin surface and the actual value RT. This measuring behavior can mislead inexperienced therapists with inaccurate values that suggest thinner dentin than the reality. In this study, a new electrode based on the technology of active guard drive was constructed to overcome this issue. The results show that improvement in the measuring behavior of the new electrode could be achieved, eliminating the earlier disadvantage of the Prepometer.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Condutometria
/
Polimento Dentário
/
Polpa Dentária
/
Dentina
/
Eletrodos
Tipo de estudo:
Diagnostic_studies
Limite:
Humans
Idioma:
En
Revista:
Biomed Tech (Berl)
Ano de publicação:
2013
Tipo de documento:
Article
País de publicação:
Alemanha