Cardiac perforation and lead dislodgement after implantation of a MR-conditional pacing lead: a single-center experience.
Elmouchi, Darryl A; Rosema, Shelly; Vanoosterhout, Stacie M; Khan, Mohsin; Davis, Alan T; Gauri, Andre J; Finta, Bohuslav; Woelfel, Alan K; Chalfoun, Nagib T.
Afiliação
  • Elmouchi DA; Frederik Meijer Heart and Vascular Institute, Spectrum Health, Grand Rapids, Michigan; Department of Internal Medicine, Michigan State University College of Human Medicine, Grand Rapids, Michigan.
Pacing Clin Electrophysiol ; 37(1): 4-10, 2014 Jan.
Article em En | MEDLINE | ID: mdl-24215291
ABSTRACT

OBJECTIVE:

The purpose of our study was to determine if the rate of lead-related complications was increased with the Medtronic CapSureFix MRI™ SureScan™ 5086 MRI pacing lead (5086; Medtronic Inc., Minneapolis, MN, USA) compared to the previous generation of Medtronic CapSureFix Novus™ 5076 pacing lead (5076).

BACKGROUND:

The 5086 lead is a newly introduced active-fixation pacemaker lead designed to be used conditionally in a magnetic resonance (MR) scanner. This lead has specific design changes compared to the previous generation of 5076 pacing leads.

METHODS:

This study was a retrospective case control study of 65 consecutive patients implanted with two 5086 leads compared to 92 consecutive control patients implanted with two 5076 leads over a 14-month period at a high-volume tertiary care hospital.

RESULTS:

Pericarditis, pericardial effusion, cardiac tamponade, or death within 30 days of implant were seen in eight patients from the 5086 cohort and two from the 5076 cohort (odds ratio 6.3, 95% confidence interval 1.3-30.8, P = 0.02). Lead dislodgement occurred in four of the 5086 patients and in none of the 5076 patients (P < 0.03).

CONCLUSIONS:

In a high-volume center, the incidence of pericarditis, cardiac perforation, tamponade, death, and lead dislodgement was significantly higher with the MR-conditional Medtronic 5086 lead when compared to the previous generation Medtronic 5076 lead.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Ferimentos Penetrantes / Imageamento por Ressonância Magnética / Migração de Corpo Estranho / Eletrodos Implantados / Traumatismos Cardíacos Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Ferimentos Penetrantes / Imageamento por Ressonância Magnética / Migração de Corpo Estranho / Eletrodos Implantados / Traumatismos Cardíacos Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2014 Tipo de documento: Article