Control time reduction using virtual source projection for treating a leg sarcoma with nonlinear perfusion.
Cheng, Kung-Shan; Yuan, Yu; Li, Zhen; Stauffer, Paul R; Joines, William T; Dewhirst, Mark W; Das, Shiva K.
Afiliação
  • Cheng KS; Division of Radiation Oncology, Duke University Medical Center, Durham, NC, USA 27710.
  • Yuan Y; Division of Radiation Oncology, Duke University Medical Center, Durham, NC, USA 27710.
  • Li Z; Department of Electric Engineering, Duke University, Durham, NC, USA 27710.
  • Stauffer PR; Division of Radiation Oncology, Duke University Medical Center, Durham, NC, USA 27710.
  • Joines WT; Department of Electric Engineering, Duke University, Durham, NC, USA 27710.
  • Dewhirst MW; Division of Radiation Oncology, Duke University Medical Center, Durham, NC, USA 27710.
  • Das SK; Division of Radiation Oncology, Duke University Medical Center, Durham, NC, USA 27710.
Proc SPIE Int Soc Opt Eng ; 71812009 Feb 12.
Article em En | MEDLINE | ID: mdl-24392195
ABSTRACT

PURPOSE:

Blood perfusion is a well-known factor that complicates accurate control of heating during hyperthermia treatments of cancer. Since blood perfusion varies as a function of time, temperature and location, determination of appropriate power deposition pattern from multiple antenna array Hyperthermia systems and heterogeneous tissues is a difficult control problem. Therefore, we investigate the applicability of a real-time eigenvalue model reduction (virtual source - VS) reduced-order controller for hyperthermic treatments of tissue with nonlinearly varying perfusion.

METHODS:

We impose a piecewise linear approximation to a set of heat pulses, each consisting of a 1-min heat-up, followed by a 2-min cool-down. The controller is designed for feedback from magnetic resonance temperature images (MRTI) obtained after each iteration of heat pulses to adjust the projected optimal setting of antenna phase and magnitude for selective tumor heating. Simulated temperature patterns with additive Gaussian noise with a standard deviation of 1.0°C and zero mean were used as a surrogate for MRTI. Robustness tests were conducted numerically for a patient's right leg placed at the middle of a water bolus surrounded by a 10-antenna applicator driven at 150 MHz. Robustness tests included added discrepancies in perfusion, electrical and thermal properties, and patient model simplifications.

RESULTS:

The controller improved selective tumor heating after an average of 4-9 iterative adjustments of power and phase, and fulfilled satisfactory therapeutic outcomes with approximately 75% of tumor volumes heated to temperatures >43°C while maintaining about 93% of healthy tissue volume < 41°C. Adequate sarcoma heating was realized by using only 2 to 3 VSs rather than a much larger number of control signals for all 10 antennas, which reduced the convergence time to only 4 to 9% of the original value.

CONCLUSIONS:

Using a piecewise linear approximation to a set of heat pulses in a VS reduced-order controller, the proposed algorithm greatly improves the efficiency of hyperthermic treatment of leg sarcomas while accommodating practical nonlinear variation of tissue properties such as perfusion.