INTERCAL: long-term inter-comparison experiment for dose rate and spectrometric probes.
Radiat Prot Dosimetry
; 160(4): 306-10, 2014 Aug.
Article
em En
| MEDLINE
| ID: mdl-24639590
ABSTRACT
The Schauinsland inter-calibration facility (INTERCAL) has been designed to enable long-term comparison experiments for 20 different dose rate probes from different networks. Two reference probes characterised by the European Radiation Dosimetry WG3 inter-calibration experiments in 2008 and 2009 have been installed at the INTERCAL facility. Additional instrumentation provides measured data of activity in air and nuclide-specific dose rate as well as environmental parameters such as air pressure, temperature, precipitation and soil moisture. Complementary to WG3 experiments, the INTERCAL platform is an ideal framework to investigate the long-term behaviour of dose rate probes and different spectrometry systems under environmental conditions. Two additional exposure experiments were performed in April 2009 and in May 2012.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Poluentes Radioativos
/
Espectrometria gama
/
Monitoramento de Radiação
/
Poluição Ambiental
Aspecto:
Determinantes_sociais_saude
Limite:
Humans
País/Região como assunto:
Europa
Idioma:
En
Revista:
Radiat Prot Dosimetry
Ano de publicação:
2014
Tipo de documento:
Article
País de afiliação:
Alemanha