Stent intussusception after thromboaspiration through a platinum chrome stent: a particular case of longitudinal stent deformation.
Mila, Rafael; Vignolo, Gustavo; Trujillo, Pedro.
Afiliação
  • Mila R; Servicio de Hemodinamia, Cátedra de Cardiología, Centro Cardiovascular Universitario, Hospital de Clínicas, Universidad de la República, Avda Italia S/N, 11600, Montevideo, Uruguay, rmila@hc.edu.uy.
Cardiovasc Interv Ther ; 30(2): 147-50, 2015 Apr.
Article em En | MEDLINE | ID: mdl-24729027
The need to improve stent deliverability has led to the development of thinner and more flexible stents. However, there is concern about decreased longitudinal strength. The number of longitudinal stent deformation reports has dramatically increased. We report a case of stent longitudinal deformation after thromboaspiration through a new generation platinum chrome bare metal stent. Images show an "intussusception effect," an extreme form of the previously described "concertina deformation," as the mechanism of shortening. Since stent technology is constantly evolving, newer devices will probably be designed to have less susceptibility to longitudinal stent deformation.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Stents / Trombectomia Limite: Aged / Female / Humans Idioma: En Revista: Cardiovasc Interv Ther Ano de publicação: 2015 Tipo de documento: Article País de publicação: Japão