Impact of reversed phase column pairs in comprehensive two-dimensional liquid chromatography.
Allen, Robert C; Barnes, Brian B; Haidar Ahmad, Imad A; Filgueira, Marcelo R; Carr, Peter W.
Afiliação
  • Allen RC; University of Minnesota, Department of Chemistry, Smith and Kolthoff Halls, 207 Pleasant Street SE, Minneapolis, MN 55455, USA.
  • Barnes BB; University of Minnesota, Department of Chemistry, Smith and Kolthoff Halls, 207 Pleasant Street SE, Minneapolis, MN 55455, USA. Electronic address: bbbarnes@umn.edu.
  • Haidar Ahmad IA; University of Minnesota, Department of Chemistry, Smith and Kolthoff Halls, 207 Pleasant Street SE, Minneapolis, MN 55455, USA.
  • Filgueira MR; University of Minnesota, Department of Chemistry, Smith and Kolthoff Halls, 207 Pleasant Street SE, Minneapolis, MN 55455, USA.
  • Carr PW; University of Minnesota, Department of Chemistry, Smith and Kolthoff Halls, 207 Pleasant Street SE, Minneapolis, MN 55455, USA.
J Chromatogr A ; 1361: 169-77, 2014 Sep 26.
Article em En | MEDLINE | ID: mdl-25169724
A major issue in optimizing the resolving power of two-dimensional chromatographic separations is the choice of the two phases so as to maximize the distribution of the analytes over the separation space. In this work, we studied the choice of appropriate reversed phases to use in on-line comprehensive two-dimensional liquid chromatography (LC×LC). A set of four chemically different conventional bonded reversed phases was used in the first dimension. The second dimension column was either a conventional bonded C18 phase or a carbon-clad phase (CCP). The LC×LC chromatograms and contour plots were all rather similar indicating that the selectivities of the two phases were also similar regardless of the reverse phase column used in the first dimension. Further, the spatial coverage seen with all four first dimension stationary phases when paired with a second dimension C18 phase were low and the retention times were strongly correlated. However, when the C18 column was replaced with the CCP column much improved separations were observed with higher spatial coverages, greater orthogonalities and significant increases in the number of observed peaks.