Your browser doesn't support javascript.
loading
Optimizing radiation dose by using advanced modelled iterative reconstruction in high-pitch coronary CT angiography.
Gordic, Sonja; Desbiolles, Lotus; Sedlmair, Martin; Manka, Robert; Plass, André; Schmidt, Bernhard; Husarik, Daniela B; Maisano, Francesco; Wildermuth, Simon; Alkadhi, Hatem; Leschka, Sebastian.
Afiliação
  • Gordic S; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Raemistrasse 100, CH-8091, Zurich, Switzerland.
  • Desbiolles L; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Raemistrasse 100, CH-8091, Zurich, Switzerland.
  • Sedlmair M; Divison of Radiology and Nuclear Medicine, Kantonsspital, St. Gallen, Switzerland.
  • Manka R; Siemens Healthcare, Computed Tomography Division, Forchheim, Germany.
  • Plass A; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Raemistrasse 100, CH-8091, Zurich, Switzerland.
  • Schmidt B; Clinic of Cardiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
  • Husarik DB; Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
  • Maisano F; Clinic for Cardiovascular Surgery, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
  • Wildermuth S; Siemens Healthcare, Computed Tomography Division, Forchheim, Germany.
  • Alkadhi H; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Raemistrasse 100, CH-8091, Zurich, Switzerland.
  • Leschka S; Clinic for Cardiovascular Surgery, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
Eur Radiol ; 26(2): 459-68, 2016 Feb.
Article em En | MEDLINE | ID: mdl-26037718
ABSTRACT

PURPOSE:

To evaluate the potential of advanced modeled iterative reconstruction (ADMIRE) for optimizing radiation dose of high-pitch coronary CT angiography (CCTA).

METHODS:

High-pitch 192-slice dual-source CCTA was performed in 25 patients (group 1) according to standard settings (ref. 100 kVp, ref. 270 mAs/rot). Images were reconstructed with filtered back projection (FBP) and ADMIRE (strength levels 1-5). In another 25 patients (group 2), high-pitch CCTA protocol parameters were adapted according to results from group 1 (ref. 160 mAs/rot), and images were reconstructed with ADMIRE level 4. In ten patients of group 1, vessel sharpness using full width at half maximum (FWHM) analysis was determined. Image quality was assessed by two independent, blinded readers.

RESULTS:

Interobserver agreements for attenuation and noise were excellent (r = 0.88/0.85, p < 0.01). In group 1, ADMIRE level 4 images were most often selected (84%, 21/25) as preferred data set; at this level noise reduction was 40% compared to FBP. Vessel borders showed increasing sharpness (FWHM) at increasing ADMIRE levels (p < 0.05). Image quality in group 2 was similar to that of group 1 at ADMIRE levels 2-3. Radiation dose in group 2 (0.3 ± 0.1 mSv) was significantly lower than in group 1 (0.5 ± 0.3 mSv; p < 0.05).

CONCLUSIONS:

In a selected population, ADMIRE can be used for optimizing high-pitch CCTA to an effective dose of 0.3 mSv. KEY POINTS • Advanced modeled IR (ADMIRE) reduces image noise up to 50% as compared to FBP. • Coronary artery vessel borders show an increasing sharpness at higher ADMIRE levels. • High-pitch CCTA with ADMIRE is possible at a radiation dose of 0.3 mSv.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doses de Radiação / Processamento de Imagem Assistida por Computador / Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X / Angiografia Coronária Tipo de estudo: Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Suíça

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doses de Radiação / Processamento de Imagem Assistida por Computador / Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X / Angiografia Coronária Tipo de estudo: Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Suíça