Your browser doesn't support javascript.
loading
Stop the Noise: A Quality Improvement Project to Decrease Electrocardiographic Nuisance Alarms.
Sendelbach, Sue; Wahl, Sharon; Anthony, Anita; Shotts, Pam.
Afiliação
  • Sendelbach S; Sue Sendelbach is director of nursing research at Abbott Northwestern Hospital, Minneapolis, Minnesota, and is on the Association of the Advancement of Medical Instrumentation's committee on alarm management/alarm fatigue.Sharon Wahl is a cardiovascular clinical nurse specialist at Abbott Northweste
  • Wahl S; Sue Sendelbach is director of nursing research at Abbott Northwestern Hospital, Minneapolis, Minnesota, and is on the Association of the Advancement of Medical Instrumentation's committee on alarm management/alarm fatigue.Sharon Wahl is a cardiovascular clinical nurse specialist at Abbott Northweste
  • Anthony A; Sue Sendelbach is director of nursing research at Abbott Northwestern Hospital, Minneapolis, Minnesota, and is on the Association of the Advancement of Medical Instrumentation's committee on alarm management/alarm fatigue.Sharon Wahl is a cardiovascular clinical nurse specialist at Abbott Northweste
  • Shotts P; Sue Sendelbach is director of nursing research at Abbott Northwestern Hospital, Minneapolis, Minnesota, and is on the Association of the Advancement of Medical Instrumentation's committee on alarm management/alarm fatigue.Sharon Wahl is a cardiovascular clinical nurse specialist at Abbott Northweste
Crit Care Nurse ; 35(4): 15-22; quiz 1p following 22, 2015 Aug.
Article em En | MEDLINE | ID: mdl-26232798
ABSTRACT

BACKGROUND:

As many as 99% of alarm signals may not need any intervention and can result in patients' deaths. Alarm management is now a Joint Commission National Patient Safety Goal.

OBJECTIVES:

To reduce the number of nuisance electrocardiographic alarm signals in adult patients on the medical cardiovascular care unit.

METHODS:

A quality improvement process was used that included eliminating duplicative alarms, customizing alarms, changing electrocardiography electrodes daily, standardizing skin preparation, and using disposable electrocardiography leads.

RESULTS:

In the cardiovascular care unit, the mean number of electrocardiographic alarm signals per day decreased from 28.5 (baseline) to 3.29, an 88.5% reduction.

CONCLUSION:

Use of a bundled approach to managing alarm signals decreased the mean number of alarm signals in a cardiovascular care unit.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Eletrocardiografia / Falha de Equipamento / Alarmes Clínicos / Melhoria de Qualidade / Segurança do Paciente / Pacotes de Assistência ao Paciente / Enfermagem de Cuidados Críticos Limite: Adult / Humans Idioma: En Revista: Crit Care Nurse Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Eletrocardiografia / Falha de Equipamento / Alarmes Clínicos / Melhoria de Qualidade / Segurança do Paciente / Pacotes de Assistência ao Paciente / Enfermagem de Cuidados Críticos Limite: Adult / Humans Idioma: En Revista: Crit Care Nurse Ano de publicação: 2015 Tipo de documento: Article