Advanced image-based virtual monoenergetic dual-energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast.
Albrecht, Moritz H; Scholtz, Jan-Erik; Hüsers, Kristina; Beeres, Martin; Bucher, Andreas M; Kaup, Moritz; Martin, Simon S; Fischer, Sebastian; Bodelle, Boris; Bauer, Ralf W; Lehnert, Thomas; Vogl, Thomas J; Wichmann, Julian L.
Afiliação
  • Albrecht MH; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany. MoritzAlbrecht@gmx.net.
  • Scholtz JE; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
  • Hüsers K; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
  • Beeres M; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
  • Bucher AM; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
  • Kaup M; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
  • Martin SS; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
  • Fischer S; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
  • Bodelle B; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
  • Bauer RW; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
  • Lehnert T; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
  • Vogl TJ; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
  • Wichmann JL; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
Eur Radiol ; 26(6): 1863-70, 2016 Jun.
Article em En | MEDLINE | ID: mdl-26334508
ABSTRACT

OBJECTIVES:

To compare quantitative image quality parameters in abdominal dual-energy computed tomography angiography (DE-CTA) using an advanced image-based (Mono+) reconstruction algorithm for virtual monoenergetic imaging and standard DE-CTA.

METHODS:

Fifty-five patients (36 men; mean age, 64.2 ± 12.7 years) who underwent abdominal DE-CTA were retrospectively included. Mono + images were reconstructed at 40, 50, 60, 70, 80, 90 and 100 keV levels and as standard linearly blended M_0.6 images (60 % 100 kV, 40 % 140 kV). The contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) of the common hepatic (CHA), splenic (SA), superior mesenteric (SMA) and left renal arteries (LRA) were objectively measured.

RESULTS:

Mono+ DE-CTA series showed a statistically superior CNR for 40, 50, 60, 70 and 80 keV (P < 0.031) compared to M_0.6 images for all investigated arteries except SMA at 80 keV (P = 0.08). CNR at 40 keV revealed a mean relative increase of 287.7 % compared to linearly blended images among all assessed arteries (P < 0.001). SNR of Mono+ images was consistently significantly higher at 40, 50, 60 and 70 keV compared to M_0.6 for CHA and SA (P < 0.009).

CONCLUSIONS:

Compared to linearly blended images, Mono+ reconstructions at low keV levels of abdominal DE-CTA datasets significantly improve quantitative image quality. KEY POINTS • Mono+ combines increased attenuation with reduced image noise compared to standard DE-CTA. • Mono+ shows superior contrast-to-noise ratios at low keV compared to linearly-blended images. • Contrast-to-noise ratio in monoenergetic DE-CTA peaks at 40 keV. • Mono+ reconstructions significantly improve quantitative image quality at low keV levels.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Imagem Radiográfica a Partir de Emissão de Duplo Fóton / Cavidade Abdominal / Angiografia por Tomografia Computadorizada Tipo de estudo: Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Imagem Radiográfica a Partir de Emissão de Duplo Fóton / Cavidade Abdominal / Angiografia por Tomografia Computadorizada Tipo de estudo: Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Alemanha