Your browser doesn't support javascript.
loading
Performance of the GeneXpert Ebola Assay for Diagnosis of Ebola Virus Disease in Sierra Leone: A Field Evaluation Study.
Semper, Amanda E; Broadhurst, M Jana; Richards, Jade; Foster, Geraldine M; Simpson, Andrew J H; Logue, Christopher H; Kelly, J Daniel; Miller, Ann; Brooks, Tim J G; Murray, Megan; Pollock, Nira R.
Afiliação
  • Semper AE; Public Health England, Porton Down, United Kingdom.
  • Broadhurst MJ; Partners In Health, Boston, Massachusetts, United States of America.
  • Richards J; Public Health England Laboratory, Port Loko, Sierra Leone.
  • Foster GM; Mid Essex Hospital Services NHS Trust, Chelmsford, United Kingdom.
  • Simpson AJ; Public Health England Laboratory, Port Loko, Sierra Leone.
  • Logue CH; Liverpool School of Tropical Medicine, Liverpool, United Kingdom.
  • Kelly JD; Public Health England, Porton Down, United Kingdom.
  • Miller A; Public Health England Laboratory, Port Loko, Sierra Leone.
  • Brooks TJ; Public Health England, Porton Down, United Kingdom.
  • Murray M; Wellbody Alliance, Freetown, Sierra Leone.
  • Pollock NR; Partners In Health, Boston, Massachusetts, United States of America.
PLoS Med ; 13(3): e1001980, 2016 Mar.
Article em En | MEDLINE | ID: mdl-27023868
ABSTRACT

BACKGROUND:

Throughout the Ebola virus disease (EVD) epidemic in West Africa, field laboratory testing for EVD has relied on complex, multi-step real-time reverse transcription PCR (RT-PCR) assays; an accurate sample-to-answer RT-PCR test would reduce time to results and potentially increase access to testing. We evaluated the performance of the Cepheid GeneXpert Ebola assay on clinical venipuncture whole blood (WB) and buccal swab (BS) specimens submitted to a field biocontainment laboratory in Sierra Leone for routine EVD testing by RT-PCR ("Trombley assay"). METHODS AND

FINDINGS:

This study was conducted in the Public Health England EVD diagnostic laboratory in Port Loko, Sierra Leone, using residual diagnostic specimens remaining after clinical testing. EDTA-WB specimens (n = 218) were collected from suspected or confirmed EVD patients between April 1 and July 20, 2015. BS specimens (n = 71) were collected as part of a national postmortem screening program between March 7 and July 20, 2015. EDTA-WB and BS specimens were tested with Xpert (targets glycoprotein [GP] and nucleoprotein [NP] genes) and Trombley (target NP gene) assays in parallel. All WB specimens were fresh; 84/218 were tested in duplicate on Xpert to compare WB sampling methods (pipette versus swab); 43/71 BS specimens had been previously frozen. In all, 7/218 (3.2%) WB and 7/71 (9.9%) BS samples had Xpert results that were reported as "invalid" or "error" and were excluded, leaving 211 WB and 64 BS samples with valid Trombley and Xpert results. For WB, 22/22 Trombley-positive samples were Xpert-positive (sensitivity 100%, 95% CI 84.6%-100%), and 181/189 Trombley-negative samples were Xpert-negative (specificity 95.8%, 95% confidence interval (CI) 91.8%-98.2%). Seven of the eight Trombley-negative, Xpert-positive (Xpert cycle threshold [Ct] range 37.7-43.4) WB samples were confirmed to be follow-up submissions from previously Trombley-positive EVD patients, suggesting a revised Xpert specificity of 99.5% (95% CI 97.0%-100%). For Xpert-positive WB samples (n = 22), Xpert NP Ct values were consistently lower than GP Ct values (mean difference -4.06, 95% limits of agreement -6.09, -2.03); Trombley (NP) Ct values closely matched Xpert NP Ct values (mean difference -0.04, 95% limits of agreement -2.93, 2.84). Xpert results (positive/negative) for WB sampled by pipette versus swab were concordant for 78/79 (98.7%) WB samples, with comparable Ct values for positive results. For BS specimens, 20/20 Trombley-positive samples were Xpert-positive (sensitivity 100%, 95% CI 83.2%-100%), and 44/44 Trombley-negative samples were Xpert-negative (specificity 100%, 95% CI 92.0%-100%). This study was limited to testing residual diagnostic samples, some of which had been frozen before use; it was not possible to test the performance of the Xpert Ebola assay at point of care.

CONCLUSIONS:

The Xpert Ebola assay had excellent performance compared to an established RT-PCR benchmark on WB and BS samples in a field laboratory setting. Future studies should evaluate feasibility and performance outside of a biocontainment laboratory setting to facilitate expanded access to testing.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: RNA Viral / Glicoproteínas / Doença pelo Vírus Ebola / Reação em Cadeia da Polimerase Via Transcriptase Reversa / Ebolavirus / Reação em Cadeia da Polimerase em Tempo Real / Nucleoproteínas Tipo de estudo: Diagnostic_studies Limite: Adolescent / Adult / Aged / Aged80 / Child / Child, preschool / Female / Humans / Infant / Male País/Região como assunto: Africa Idioma: En Revista: PLoS Med Assunto da revista: MEDICINA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: RNA Viral / Glicoproteínas / Doença pelo Vírus Ebola / Reação em Cadeia da Polimerase Via Transcriptase Reversa / Ebolavirus / Reação em Cadeia da Polimerase em Tempo Real / Nucleoproteínas Tipo de estudo: Diagnostic_studies Limite: Adolescent / Adult / Aged / Aged80 / Child / Child, preschool / Female / Humans / Infant / Male País/Região como assunto: Africa Idioma: En Revista: PLoS Med Assunto da revista: MEDICINA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Reino Unido