Dose controlled low energy electron irradiator for biomolecular films.
Rev Sci Instrum
; 87(3): 034302, 2016 Mar.
Article
em En
| MEDLINE
| ID: mdl-27036792
We have developed a multi target, Low Energy Electron (LEE), precise dose controlled irradiator for biomolecular films. Up to seven samples can be irradiated one after another at any preset electron energy and dose under UHV conditions without venting the chamber. In addition, one more sample goes through all the steps except irradiation, which can be used as control for comparison with the irradiated samples. All the samples are protected against stray electron irradiation by biasing them at -20 V during the entire period, except during irradiation. Ethernet based communication electronics hardware, LEE beam control electronics and computer interface were developed in house. The user Graphical User Interface to control the irradiation and dose measurement was developed using National Instruments Lab Windows CVI. The working and reliability of the dose controlled irradiator has been fully tested over the electron energy range of 0.5 to 500 eV by studying LEE induced single strand breaks to ΦX174 RF1 dsDNA.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Doses de Radiação
/
DNA
/
Elétrons
/
Equipamentos e Provisões para Radiação
Idioma:
En
Revista:
Rev Sci Instrum
Ano de publicação:
2016
Tipo de documento:
Article
País de afiliação:
Índia
País de publicação:
Estados Unidos