Validation of modelled imaging plates sensitivity to 1-100 keV x-rays and spatial resolution characterisation for diagnostics for the "PETawatt Aquitaine Laser".
Boutoux, G; Batani, D; Burgy, F; Ducret, J-E; Forestier-Colleoni, P; Hulin, S; Rabhi, N; Duval, A; Lecherbourg, L; Reverdin, C; Jakubowska, K; Szabo, C I; Bastiani-Ceccotti, S; Consoli, F; Curcio, A; De Angelis, R; Ingenito, F; Baggio, J; Raffestin, D.
Afiliação
  • Boutoux G; Université de Bordeaux, CNRS, CEA, CELIA (Centre Lasers Intenses et Applications), UMR 5107, F-33405 Talence, France.
  • Batani D; Université de Bordeaux, CNRS, CEA, CELIA (Centre Lasers Intenses et Applications), UMR 5107, F-33405 Talence, France.
  • Burgy F; Université de Bordeaux, CNRS, CEA, CELIA (Centre Lasers Intenses et Applications), UMR 5107, F-33405 Talence, France.
  • Ducret JE; Université de Bordeaux, CNRS, CEA, CELIA (Centre Lasers Intenses et Applications), UMR 5107, F-33405 Talence, France.
  • Forestier-Colleoni P; Université de Bordeaux, CNRS, CEA, CELIA (Centre Lasers Intenses et Applications), UMR 5107, F-33405 Talence, France.
  • Hulin S; Université de Bordeaux, CNRS, CEA, CELIA (Centre Lasers Intenses et Applications), UMR 5107, F-33405 Talence, France.
  • Rabhi N; Université de Bordeaux, CNRS, CEA, CELIA (Centre Lasers Intenses et Applications), UMR 5107, F-33405 Talence, France.
  • Duval A; CEA DAM DIF, F-91297 Arpajon, France.
  • Lecherbourg L; CEA DAM DIF, F-91297 Arpajon, France.
  • Reverdin C; CEA DAM DIF, F-91297 Arpajon, France.
  • Jakubowska K; Institute of Plasma Physics and Laser Microfusion, Hery Street 23, 01-497 Warsaw, Poland.
  • Szabo CI; Theiss Research, 7411 Eads Avenue, La Jolla, California 92037, USA.
  • Bastiani-Ceccotti S; LULI, Ecole Polytechnique, CNRS, CEA, UPMC, 91128 Palaiseau, France.
  • Consoli F; ENEA for EUROfusion, Via E. Fermi 45, 00044 Frascati, Rome, Italy.
  • Curcio A; ENEA for EUROfusion, Via E. Fermi 45, 00044 Frascati, Rome, Italy.
  • De Angelis R; ENEA for EUROfusion, Via E. Fermi 45, 00044 Frascati, Rome, Italy.
  • Ingenito F; ENEA for EUROfusion, Via E. Fermi 45, 00044 Frascati, Rome, Italy.
  • Baggio J; CEA DAM CESTA, BP 12, F-33405 Le Barp, France.
  • Raffestin D; CEA DAM CESTA, BP 12, F-33405 Le Barp, France.
Rev Sci Instrum ; 87(4): 043108, 2016 04.
Article em En | MEDLINE | ID: mdl-27131655
ABSTRACT
Thanks to their high dynamic range and ability to withstand electromagnetic pulse, imaging plates (IPs) are commonly used as passive detectors in laser-plasma experiments. In the framework of the development of the diagnostics for the Petawatt Aquitaine Laser facility, we present an absolute calibration and spatial resolution study of five different available types of IP (namely, MS-SR-TR-MP-ND) performed by using laser-induced K-shell X-rays emitted by a solid silver target irradiated by the laser ECLIPSE at CEntre Lasers Intenses et Applications. In addition, IP sensitivity measurements were performed with a 160 kV X-ray generator at CEA DAM DIF, where the absolute response of IP SR and TR has been calibrated to X-rays in the energy range 8-75 keV with uncertainties of about 15%. Finally, the response functions have been modeled in Monte Carlo GEANT4 simulations in order to reproduce experimental data. Simulations enable extrapolation of the IP response functions to photon energies from 1 keV to 1 GeV, of interest, e.g., for laser-driven radiography.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2016 Tipo de documento: Article País de afiliação: França