Evaluation of a Portable Artificial Vision Device Among Patients With Low Vision.
Moisseiev, Elad; Mannis, Mark J.
Afiliação
  • Moisseiev E; Department of Ophthalmology and Vision Science, University of California Davis Eye Center, Sacramento.
  • Mannis MJ; Department of Ophthalmology and Vision Science, University of California Davis Eye Center, Sacramento.
JAMA Ophthalmol ; 134(7): 748-52, 2016 Jul 01.
Article em En | MEDLINE | ID: mdl-27148909
ABSTRACT
IMPORTANCE Low vision is irreversible in many patients and constitutes a disability. When no treatment to improve vision is available, technological developments aid these patients in their daily lives.

OBJECTIVE:

To evaluate the usefulness of a portable artificial vision device (OrCam) for patients with low vision. DESIGN, SETTING, AND

PARTICIPANTS:

A prospective pilot study was conducted between July 1 and September 30, 2015, in a US ophthalmology department among 12 patients with visual impairment and best-corrected visual acuity of 20/200 or worse in their better eye.

INTERVENTIONS:

A 10-item test simulating activities of daily living was used to evaluate patients' functionality in 3 scenarios using their best-corrected visual acuity with no low-vision aids, using low-vision aids if available, and using the portable artificial vision device. This 10-item test was devised for this study and is nonvalidated. The portable artificial vision device was tested at the patients' first visit and after 1 week of use at home. MAIN OUTCOMES AND

MEASURES:

Scores on the 10-item daily function test.

RESULTS:

Among the 12 patients, scores on the 10-item test improved from a mean (SD) of 2.5 (1.6) using best-corrected visual acuity to 9.5 (0.5) using the portable artificial vision device at the first visit (mean difference, 7.0; 95% CI, 6.0-8.0; P < .001) and 9.8 (0.4) after 1 week (mean difference from the first visit, 7.3; 95% CI, 6.3-8.3; P < .001). Mean (SD) scores with the portable artificial vision device were also better in the 7 patients who used other low-vision aids (9.7 [0.5] vs 6.0 [2.6], respectively; mean difference, 3.7; 95% CI, 1.5-5.9; P = .01). CONCLUSIONS AND RELEVANCE When patients used a portable artificial vision device, an increase in scores on a nonvalidated 10-item test of activities of daily living was seen. Further evaluations are warranted to determine the usefulness of this device among individuals with low vision.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Auxiliares Sensoriais / Fotografação / Baixa Visão Tipo de estudo: Evaluation_studies / Observational_studies Aspecto: Patient_preference Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: JAMA Ophthalmol Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Auxiliares Sensoriais / Fotografação / Baixa Visão Tipo de estudo: Evaluation_studies / Observational_studies Aspecto: Patient_preference Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: JAMA Ophthalmol Ano de publicação: 2016 Tipo de documento: Article