Your browser doesn't support javascript.
loading
The prediction of drop size from intravenous infusion controllers.
Hillman, M R.
Afiliação
  • Hillman MR; Bath Institute of Medical Engineering, Wolfson Centre, Royal United Hospital, Combe Park, U.K.
J Med Eng Technol ; 13(3): 166-76, 1989.
Article em En | MEDLINE | ID: mdl-2746624
In intravenous therapy, gravity-fed drip controllers regulate the flow of fluid in terms of drops per minute, though the total amount of fluid to be infused is specified in ml/h. To convert from one to the other it is regularly, erroneously, assumed that drop size is constant. The work described here investigates the relationship of drop size with respect to drip rate and fluid properties. The results are presented in an easy-to-use graphical form.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Infusões Intravenosas Tipo de estudo: Prognostic_studies / Risk_factors_studies Idioma: En Revista: J Med Eng Technol Ano de publicação: 1989 Tipo de documento: Article País de publicação: Reino Unido
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Infusões Intravenosas Tipo de estudo: Prognostic_studies / Risk_factors_studies Idioma: En Revista: J Med Eng Technol Ano de publicação: 1989 Tipo de documento: Article País de publicação: Reino Unido