Your browser doesn't support javascript.
loading
A multiplexed TOF and DOI capable PET detector using a binary position sensitive network.
Bieniosek, M F; Cates, J W; Levin, C S.
Afiliação
  • Bieniosek MF; Department of Radiology, Stanford University, Stanford, CA 94305, USA. Molecular Imaging Program at Stanford (MIPS), Stanford, CA 94305, USA. Department of Electrical Engineering, Stanford University, Stanford, CA 94305, USA.
Phys Med Biol ; 61(21): 7639-7651, 2016 11 07.
Article em En | MEDLINE | ID: mdl-27740946
Time of flight (TOF) and depth of interaction (DOI) capabilities can significantly enhance the quality and uniformity of positron emission tomography (PET) images. Many proposed TOF/DOI PET detectors require complex readout systems using additional photosensors, active cooling, or waveform sampling. This work describes a high performance, low complexity, room temperature TOF/DOI PET module. The module uses multiplexed timing channels to significantly reduce the electronic readout complexity of the PET detector while maintaining excellent timing, energy, and position resolution. DOI was determined using a two layer light sharing scintillation crystal array with a novel binary position sensitive network. A 20 mm effective thickness LYSO crystal array with four 3 mm × 3 mm silicon photomultipliers (SiPM) read out by a single timing channel, one energy channel and two position channels achieved a full width half maximum (FWHM) coincidence time resolution of 180 ± 2 ps with 10 mm of DOI resolution and 11% energy resolution. With sixteen 3 mm × 3 mm SiPMs read out by a single timing channel, one energy channel and four position channels a coincidence time resolution 204 ± 1 ps was achieved with 10 mm of DOI resolution and 15% energy resolution. The methods presented here could significantly simplify the construction of high performance TOF/DOI PET detectors.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Silício / Tomografia por Emissão de Pósitrons Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Phys Med Biol Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Silício / Tomografia por Emissão de Pósitrons Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Phys Med Biol Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Reino Unido