Your browser doesn't support javascript.
loading
Decreased patient exposure to ionizing radiation during interventional rheumatology procedures after optimization of protection.
Cozic, Céline; Audran, Fabien; Blanchard, Christophe; David, Christophe; Andre, Vincent; Caulier, Michel; Varin, Stéphane; Tanguy, Gilles; Dimet, Jérôme; Cormier, Grégoire.
Afiliação
  • Cozic C; Service de rhumatologie, CHD Vendée, boulevard Stéphane-Moreau, 85925 La Roche-sur-Yon cedex 9, France.
  • Audran F; Unité de physique médicale, CHD Vendée, 85925 La Roche-sur-Yon cedex 9, France.
  • Blanchard C; Service de radiologie, CHD Vendée, 85925 La Roche-sur-Yon cedex 9, France.
  • David C; Unité de radioprotection, CHD Vendée, 85925 La Roche-sur-Yon cedex 9, France.
  • Andre V; Service de rhumatologie, CHD Vendée, boulevard Stéphane-Moreau, 85925 La Roche-sur-Yon cedex 9, France.
  • Caulier M; Service de rhumatologie, CHD Vendée, boulevard Stéphane-Moreau, 85925 La Roche-sur-Yon cedex 9, France.
  • Varin S; Service de rhumatologie, CHD Vendée, boulevard Stéphane-Moreau, 85925 La Roche-sur-Yon cedex 9, France.
  • Tanguy G; Service de rhumatologie, CHD Vendée, boulevard Stéphane-Moreau, 85925 La Roche-sur-Yon cedex 9, France.
  • Dimet J; Service de recherche clinique, CHD Vendée, 85925 La Roche-sur-Yon cedex 9, France.
  • Cormier G; Service de rhumatologie, CHD Vendée, boulevard Stéphane-Moreau, 85925 La Roche-sur-Yon cedex 9, France. Electronic address: gregoire.cormier@chd-vendee.fr.
Joint Bone Spine ; 84(3): 335-339, 2017 May.
Article em En | MEDLINE | ID: mdl-27825575
ABSTRACT

OBJECTIVES:

To decrease radiation exposure of patients undergoing interventional rheumatology procedures, without adversely affecting quality of care.

METHODS:

The radiation dose received, assessed by the dose-area product (DAP), was measured during 283 intraarticular injections performed under fluoroscopic guidance between May and July 2013. Then, three steps were taken to decrease patients' radiation exposure a copper filter was added, the anti-scatter grid was removed, and exposure cell sensitivity was set at the highest value. DAP was measured during 158 intraarticular injections performed in 2014 with these measures in place.

RESULTS:

Mean DAP before optimization was 175µGray·m2 during facet joint injections (n=4) and 43µGray·m2 during hip injections but was less than 20µGray·m2 for injections into the shoulders (15.7µGray·m2), ankles (7.7µGray·m2), wrists (3.7µGray·m2), and fingers (3.3µGray·m2). After optimization, DAP decreased markedly for all injection sites, by 52% (shoulders) to 87% (facet joints, 22.7µGray·m2). Decreases occurred at all three steps of the procedure, i.e., patient installation, injection, and last image hold. Exposure during facet joint injections varied from 84 (54.5-108.5) µGray·m2 when body mass index (BMI) was <25kg/m2 to 228.9 (161.3-340.4)µGray·m2 when BMI was>30kg/m2.

CONCLUSION:

Simple technical changes translate into large decreases in patient radiation exposure during fluoroscopically-guided injections, particularly at the facet joints and in obese patients.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doses de Radiação / Proteção Radiológica / Fluoroscopia / Radiografia Intervencionista / Exposição à Radiação / Injeções Intra-Articulares Limite: Humans Idioma: En Revista: Joint Bone Spine Assunto da revista: REUMATOLOGIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: França

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doses de Radiação / Proteção Radiológica / Fluoroscopia / Radiografia Intervencionista / Exposição à Radiação / Injeções Intra-Articulares Limite: Humans Idioma: En Revista: Joint Bone Spine Assunto da revista: REUMATOLOGIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: França