Evaluation of a new handheld point-of-care blood gas analyser using 100 equine blood samples.
Bardell, David; West, Eleanor; Mark Senior, J.
Afiliação
  • Bardell D; School of Veterinary Science, Leahurst Campus, University of Liverpool, Neston, UK. Electronic address: david.bardell@liverpool.ac.uk.
  • West E; School of Veterinary Science, Leahurst Campus, University of Liverpool, Neston, UK.
  • Mark Senior J; School of Veterinary Science, Leahurst Campus, University of Liverpool, Neston, UK.
Vet Anaesth Analg ; 44(1): 77-85, 2017 01.
Article em En | MEDLINE | ID: mdl-28237685
ABSTRACT

OBJECTIVES:

To determine whether the Enterprise point-of-care blood analysis system (EPOC) produces results in agreement with two other blood gas analysers in regular clinical use (i-STAT and Radiometer ABL77) and to investigate the precision of the new machine when used with equine whole blood. STUDY

DESIGN:

Prospective, randomized, non-blinded, comparative laboratory analyser study. ANIMALS Horses admitted to a university teaching hospital requiring arterial or venous blood gas analysis as part of their routine clinical management.

METHODS:

One hundred equine blood samples were run immediately, consecutively and in randomized order on three blood gas analysers. Results of variables common to all three analysers were tested for agreement and compared with guidelines used in human medicine. These require 80% of results from the test analyser to fall within a defined range or percentage of results from the comparator devices to achieve acceptability. Additionally, 21 samples were run twice in quick succession on the EPOC analyser to investigate precision.

RESULTS:

Agreement targets were not met for haematocrit, haemoglobin and base excess for either i-STAT or ABL77 analysers. EPOC precision targets were not met for partial pressure of carbon dioxide, ionized calcium, haematocrit and haemoglobin. Overall comparative performance of the EPOC was good to excellent for pH, oxygen tension, potassium, bicarbonate and oxygen saturation of haemoglobin, but marginal to poor for other parameters. CONCLUSIONS AND CLINICAL RELEVANCE The EPOC may be useful in performing analysis of equine whole blood, but trend analysis of carbon dioxide tension, ionized calcium, haematocrit and haemoglobin should be interpreted with caution. The EPOC should not be used interchangeably with other blood gas analysers.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Clinical_trials / Guideline Idioma: En Revista: Vet Anaesth Analg Assunto da revista: ANESTESIOLOGIA / MEDICINA VETERINARIA Ano de publicação: 2017 Tipo de documento: Article