[Comparison of refraction with or without cycloplegia using Retinomax® or Plusoptix® devices]. / Comparaison de la réfraction avec ou sans cycloplégie au Retinomax® vs Plusoptix®.
Bui Quoc, E; Guilmin Crepon, S; Tinguely, S; Lavallee, G; Busquet, G; Angot, M; Vera, L.
Afiliação
  • Bui Quoc E; Service d'ophtalmologie, hôpital universitaire Robert-Debré, Assistance publique-Hôpitaux de Paris, 48, boulevard Sérurier, 75019 Paris, France. Electronic address: emmanuelbui@outlook.com.
  • Guilmin Crepon S; Unité de recherche clinique, hôpital universitaire Robert-Debré, Assistance publique-Hôpitaux de Paris, 48, boulevard Sérurier, 75019 Paris, France.
  • Tinguely S; Service d'ophtalmologie, hôpital universitaire Robert-Debré, Assistance publique-Hôpitaux de Paris, 48, boulevard Sérurier, 75019 Paris, France.
  • Lavallee G; Service d'ophtalmologie, hôpital universitaire Robert-Debré, Assistance publique-Hôpitaux de Paris, 48, boulevard Sérurier, 75019 Paris, France.
  • Busquet G; Service d'ophtalmologie, hôpital universitaire Robert-Debré, Assistance publique-Hôpitaux de Paris, 48, boulevard Sérurier, 75019 Paris, France.
  • Angot M; Service d'ophtalmologie, hôpital universitaire Robert-Debré, Assistance publique-Hôpitaux de Paris, 48, boulevard Sérurier, 75019 Paris, France.
  • Vera L; Service d'ophtalmologie, hôpital universitaire Robert-Debré, Assistance publique-Hôpitaux de Paris, 48, boulevard Sérurier, 75019 Paris, France.
J Fr Ophtalmol ; 40(3): 209-214, 2017 Mar.
Article em Fr | MEDLINE | ID: mdl-28318725
ABSTRACT

PURPOSE:

To evaluate the refraction in children measured with Plusoptix® without cycloplegia vs. Retinomax® apparatus with cycloplegia. PATIENTS AND

METHODS:

Measure of refraction with Plusoptix® in children>1year old referred for systematic vision screening, then measurement after cycloplegia with cyclopentolate by the Retinomax® device.

RESULTS:

Thirty-three children were included, i.e. 66eyes. Mean age was 40.7months (minimum 12; maximum 114). The Spearman correlation coefficient for the spherical equivalent was 0.52 (Plusoptix® vs. Retinomax® comparison; P<0.0001=moderate correlation). The Spearman correlation coefficient was 0.73 for astigmatism (Plusoptix® vs. Retinomax® comparison; P<0.0001=strong correlation). The Plusoptix® sensitivity for measurement of refraction was 57%, 43% and 43% respectively for spherical equivalent, sphere and astigmatism. DISCUSSION AND

CONCLUSION:

The correlation of astigmatism values is strong, whereas the correlation of sphere values is moderate. Plusoptix® seems to be unable to measure the exact refraction, because there is too large a dispersion of refraction measurements with Plusoptix®, compared to the exact refraction measured with the Retinomax®. Moreover, the sensitivity of Plusoptix® is low. Cycloplegic refraction remains indispensable in children.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Refração Ocular / Erros de Refração / Seleção Visual / Distúrbios Pupilares Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Child / Child, preschool / Humans / Infant Idioma: Fr Revista: J Fr Ophtalmol Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Refração Ocular / Erros de Refração / Seleção Visual / Distúrbios Pupilares Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Child / Child, preschool / Humans / Infant Idioma: Fr Revista: J Fr Ophtalmol Ano de publicação: 2017 Tipo de documento: Article