The influence of cement thickness on stem subsidence and cement creep in a collarless polished tapered stem: When are thick cement mantles detrimental?
Takahashi, E; Kaneuji, A; Tsuda, R; Numata, Y; Ichiseki, T; Fukui, K; Kawahara, N.
Afiliação
  • Takahashi E; Department of Orthopaedic Surgery, Kanazawa Medical University, Daigaku, Uchinada-machi, Kahoku-gun, Ishikawa 920-0293, Japan.
  • Kaneuji A; Department of Orthopaedic Surgery, Kanazawa Medical University, Daigaku, Uchinada-machi, Kahoku-gun, Ishikawa 920-0293, Japan kaneuji@kanazawa-med.ac.jp.
  • Tsuda R; Department of Orthopaedic Surgery, Kanazawa Medical University, Daigaku, Uchinada-machi, Kahoku-gun, Ishikawa 920-0293, Japan.
  • Numata Y; Department of Orthopaedic Surgery, Kanazawa Medical University, Daigaku, Uchinada-machi, Kahoku-gun, Ishikawa 920-0293, Japan.
  • Ichiseki T; Department of Orthopaedic Surgery, Kanazawa Medical University, Daigaku, Uchinada-machi, Kahoku-gun, Ishikawa 920-0293, Japan.
  • Fukui K; Department of Orthopaedic Surgery, Kanazawa Medical University, Daigaku, Uchinada-machi, Kahoku-gun, Ishikawa 920-0293, Japan.
  • Kawahara N; Department of Orthopaedic Surgery, Kanazawa Medical University, Daigaku, Uchinada-machi, Kahoku-gun, Ishikawa 920-0293, Japan.
Bone Joint Res ; 6(5): 351-357, 2017 May.
Article em En | MEDLINE | ID: mdl-28566327
ABSTRACT

OBJECTIVES:

Favourable results for collarless polished tapered stems have been reported, and cement creep due to taper slip may be a contributing factor. However, the ideal cement thickness around polished stems remains unknown. We investigated the influence of cement thickness on stem subsidence and cement creep.

METHODS:

We cemented six collarless polished tapered (CPT) stems (two stems each of small, medium and large sizes) into composite femurs that had been reamed with a large CPT rasp to achieve various thicknesses of the cement mantle. Two or three tantalum balls were implanted in the proximal cement in each femur. A cyclic loading test was then performed for each stem. The migration of the balls was measured three-dimensionally, using a micro-computed tomography (CT) scanner, before and after loading. A digital displacement gauge was positioned at the stem shoulder, and stem subsidence was measured continuously by the gauge. Final stem subsidence was measured at the balls at the end of each stem.

RESULTS:

A strong positive correlation was observed between mean cement thickness and stem subsidence in the CT slices on the balls. In the small stems, the balls moved downward to almost the same extent as the stem. There was a significant negative correlation between cement thickness and the horizontaldownward ratio of ball movement.

CONCLUSION:

Collarless polished tapered stems with thicker cement mantles resulted in greater subsidence of both stem and cement. This suggests that excessive thickness of the cement mantle may interfere with effective radial cement creep.Cite this article E. Takahashi, A. Kaneuji, R. Tsuda, Y. Numata, T. Ichiseki, K. Fukui, N. Kawahara. The influence of cement thickness on stem subsidence and cement creep in a collarless polished tapered stem When are thick cement mantles detrimental? Bone Joint Res 2017;6-357. DOI 10.1302/2046-3758.65.BJR-2017-0028.R1.