Thermal diffusivity and perfusion constants from in vivo MR-guided focussed ultrasound treatments: a feasibility study.
Dillon, Christopher R; Rieke, Viola; Ghanouni, Pejman; Payne, Allison.
Afiliação
  • Dillon CR; a Department of Radiology and Imaging Sciences , University of Utah , Salt Lake City , UT , USA.
  • Rieke V; b Department of Radiology and Biomedical Imaging , University of California San Francisco , San Francisco , CA.
  • Ghanouni P; c Department of Radiology , Stanford University , Stanford , CA , USA.
  • Payne A; d Department of Radiology and Imaging Sciences , University of Utah , Salt Lake City , UT , USA.
Int J Hyperthermia ; 34(4): 352-362, 2018 06.
Article em En | MEDLINE | ID: mdl-28595499
PURPOSE: This study investigates the feasibility of non-invasively determining thermal diffusivity (α) and the Pennes perfusion parameter (w) from pre-clinical and clinical magnetic resonance-guided focussed ultrasound (MRgFUS) temperature data. MATERIALS AND METHODS: Pre-clinical MRgFUS experiments were performed in rabbit muscle (N = 3, 28 sonications) using three-dimensional MR thermometry. Eight sonications were made in a clinical QA phantom with two-dimensional thermometry. Retrospective property determination was performed on clinical uterine fibroid (N = 8, 9 sonications) and desmoid tumour (N = 4, 7 sonications) data. The property determination method fits an analytical solution to MRgFUS temperatures in the coronal MR plane, including all temperatures acquired during heating and one cooling image. When possible, additional cooling data were acquired for property determination. RESULTS: Rabbit α and w from Heating Data (α = 0.164 mm2s-1, w = 7.9 kg m-3 s-1) and Heating and Cooling Data (α = 0.146 mm2s-1, w = 3.3 kg m-3 s-1) were within the range of gold-standard invasive measurements, with >50% reduction in variability by including cooling data. QA phantom property determination with cooling data yielded properties within 3% of expected values (α = 0.144 mm2s-1, w = 0.0 kg m-3 s-1), a difference that was not statistically significant (p = 0.053). Uterine fibroid (Heating Data: α = 0.212 mm2s-1, w = 11.0 kg m-3 s-1) and desmoid tumour (Heating & Cooling Data: α = 0.245 mm2s-1, w = 4.7 kg m-3 s-1) properties are feasible but lack independent verification. CONCLUSIONS: Thermal diffusivity and the Pennes perfusion parameter can be obtained from in vivo data and with clinical MRgFUS protocols. Property values are consistently improved by including cooling data. The utility of this property determination method will increase as clinical protocols implement improved temperature imaging.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fibromatose Agressiva / Ablação por Ultrassom Focalizado de Alta Intensidade / Leiomioma Tipo de estudo: Guideline Limite: Animals / Female / Humans Idioma: En Revista: Int J Hyperthermia Assunto da revista: NEOPLASIAS / TERAPEUTICA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fibromatose Agressiva / Ablação por Ultrassom Focalizado de Alta Intensidade / Leiomioma Tipo de estudo: Guideline Limite: Animals / Female / Humans Idioma: En Revista: Int J Hyperthermia Assunto da revista: NEOPLASIAS / TERAPEUTICA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Reino Unido