Microchannel system for rate-controlled, sequential, and pH-responsive drug delivery.
Yang, Dasom; Lee, Jung Seung; Choi, Chang-Kuk; Lee, Hong-Pyo; Cho, Seung-Woo; Ryu, WonHyoung.
Afiliação
  • Yang D; Department of Mechanical Engineering, Yonsei University, 50 Yonsei-ro, Seodaemun-gu, Seoul 120-749, Republic of Korea.
  • Lee JS; Department of Biotechnology, Yonsei University, 50 Yonsei-ro, Seodaemun-gu, Seoul 120-749, Republic of Korea.
  • Choi CK; Department of Mechanical Engineering, Yonsei University, 50 Yonsei-ro, Seodaemun-gu, Seoul 120-749, Republic of Korea.
  • Lee HP; Department of Mechanical Engineering, Yonsei University, 50 Yonsei-ro, Seodaemun-gu, Seoul 120-749, Republic of Korea.
  • Cho SW; Department of Biotechnology, Yonsei University, 50 Yonsei-ro, Seodaemun-gu, Seoul 120-749, Republic of Korea.
  • Ryu W; Department of Mechanical Engineering, Yonsei University, 50 Yonsei-ro, Seodaemun-gu, Seoul 120-749, Republic of Korea. Electronic address: whryu@yonsei.ac.kr.
Acta Biomater ; 68: 249-260, 2018 03 01.
Article em En | MEDLINE | ID: mdl-29269333
ABSTRACT
Controlled delivery of drug at a constant rate, in a sequential order, or responsive to environment conditions has been pursued for a long time to enhance the efficacy of therapeutic molecules and to minimize side effects of highly potent drugs. However, achieving such delicately-controlled delivery of a drug molecule is non-trivial and still remains a challenge. We propose the use of microchannels to control the rate, sequence, and pH-responsiveness of drug delivery for high precision and predictability. In this study, we introduce elementary drug delivery units consisting of micro-reservoirs and microchannels that have variations in their lengths, widths, numbers, and straightness. The release study demonstrates that the release rates of model drugs can be modulated by the design of microchannels. Finite element modeling of drug release predicts the performance of the drug delivery units with high accuracy. The possibility of sequential drug delivery is also demonstrated using biodegradable polymer plug in microchannels. Finally, pH-responsive delivery of drugs in microfluidic units is also discussed and demonstrated via cell viability tests. STATEMENT OF

SIGNIFICANCE:

In this work, we developed microchannel-based drug delivery devices whose release rate could be accurately calculated and controlled by design of microchannel geometry. Although there have been many advances in microfabricated drug delivery systems, in particular, reservoir-based systems, no systematic investigation has been made to utilize the release channels. In our work, an equivalent electrical circuit concept was applied to the microfluidic systems for more detailed design and analysis. A microfluidic channel was regarded as an electrical resistor; their diffusion/electrical flux could be tuned with geometric factors such as length, width, a number of channel/resistor and their connections. Furthermore, from delivery rate control using channel geometry, multifunctional channel-based release systems for sequential and pH-responsive were demonstrated.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sistemas de Liberação de Medicamentos / Microfluídica Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Acta Biomater Ano de publicação: 2018 Tipo de documento: Article