Detection of microembolic signals in the common carotid artery using Doppler sonography in the porcine model of acute heart failure treated by veno-arterial extracorporeal membrane oxygenation.
Janák, D; Hála, P; Mlcek, M; Popková, M; Lacko, S; Kudlicka, J; Kittnar, O.
Afiliação
  • Janák D; Department of Physiology, First Faculty of Medicine, Charles University, Prague, Czech Republic. janakdavid@seznam.cz.
Physiol Res ; 66(Suppl 4): S529-S536, 2017 12 30.
Article em En | MEDLINE | ID: mdl-29355381
Veno-arterial extracorporeal membrane oxygenation (VA-ECMO) is a method used for the treatment most severe cases of decompensated heart failure. The purpose of this study was to evaluate the risk of the formation of microembolisms during VA-ECMO-based therapy. Heart failure was induced with simultaneous detection of microembolisms and the measurement of blood flow rate in the common carotid artery (CCA) without VA-ECMO (0 l/min) and at the VA-ECMO blood flow rate of 1, 2, 3 and 4 l/min. If embolisms for VA-ECMO 0 l/min and the individual regimes for VA-ECMO 1, 2, 3, 4 l/min are compared, a higher VA-ECMO flow rate is accompanied by a higher number of microembolisms. The final microembolism value at 16 min was for the VA-ECMO flow rate of 0 l/min 0.0 (0, 1), VA-ECMO l/min 7.5 (4, 19), VA-ECMO 2 l/min 12.5 (4, 26), VA-ECMO 3 l/min, 21.0 (18, 57) and VA-ECMO 4 l/min, 27.5 (21, 64). Such a comparison is statistically significant if VA-ECMO 0 vs. 4 l/min p<0.0001, 0 vs. 3 l/min p<0.01 and 1 vs. 4 l/min p<0.01 are compared. The results confirm that high VA-ECMO flow rates pose a risk with regards to the formation of a significantly higher number of microemboli in the blood circulation and that an increase in blood flow rates in the CCA corresponds to changes in the VA-ECMO flow rates.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxigenação por Membrana Extracorpórea / Artéria Carótida Primitiva / Ultrassonografia Doppler / Modelos Animais de Doenças / Embolia / Insuficiência Cardíaca Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Revista: Physiol Res Assunto da revista: FISIOLOGIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: República Tcheca País de publicação: República Tcheca
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxigenação por Membrana Extracorpórea / Artéria Carótida Primitiva / Ultrassonografia Doppler / Modelos Animais de Doenças / Embolia / Insuficiência Cardíaca Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Revista: Physiol Res Assunto da revista: FISIOLOGIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: República Tcheca País de publicação: República Tcheca