Ivanovic, M D; Petrovic, J; Savic, A; Gligoric, G; Miletic, M; Vukcevic, M; Bojovic, B; Hadzievski, Lj; Allsop, T; Webb, D J.
Afiliação
  • Ivanovic MD; Vinca Institute of Nuclear Sciences, University of Belgrade, Mike Petrovica Alasa 12-14, 11001 Belgrade, Serbia.
Physiol Meas ; 39(4): 045009, 2018 04 26.
Article em En | MEDLINE | ID: mdl-29553480
ABSTRACT

OBJECTIVE:

The ventilators involved in non-invasive mechanical ventilation commonly provide ventilator support via a facemask. The interface of the mask with a patient promotes air leaks that cause errors in the feedback information provided by a pneumatic sensor and hence patient-ventilator asynchrony with multiple negative consequences. Our objective is to test the possibility of using chest-wall motion measured by an optical fibre-grating sensor as a more accurate non-invasive ventilator triggering mechanism.

APPROACH:

The basic premise of our approach is that the measurement accuracy can be improved by using a triggering signal that precedes pneumatic triggering in the neuro-ventilatory coupling sequence. We propose a technique that uses the measurement of chest-wall curvature by a long-period fibre-grating sensor. The sensor was applied externally to the rib-cage and interrogated in the lateral (edge) filtering scheme. The study was performed on 34 healthy volunteers. Statistical data analysis of the time lag between the fibre-grating sensor and the reference pneumotachograph was preceded by the removal of the unwanted heartbeat signal by wavelet transform processing. MAIN

RESULTS:

The results show a consistent fibre-grating signal advance with respect to the standard pneumatic signal by (230 ± 100) ms in both the inspiratory and expiratory phases. We further show that heart activity removal yields a tremendous improvement in sensor accuracy by reducing it from 60 ml to 0.3 ml.

SIGNIFICANCE:

The results indicate that the proposed measurement technique may lead to a more reliable triggering decision. Its imperviousness to air leaks, non-invasiveness, low-cost and ease of implementation offer good prospects for applications in both clinical and homecare ventilation.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ventiladores Mecânicos / Parede Torácica / Fibras Ópticas / Movimento Tipo de estudo: Clinical_trials / Prognostic_studies Idioma: En Revista: Physiol Meas Assunto da revista: BIOFISICA / ENGENHARIA BIOMEDICA / FISIOLOGIA Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ventiladores Mecânicos / Parede Torácica / Fibras Ópticas / Movimento Tipo de estudo: Clinical_trials / Prognostic_studies Idioma: En Revista: Physiol Meas Assunto da revista: BIOFISICA / ENGENHARIA BIOMEDICA / FISIOLOGIA Ano de publicação: 2018 Tipo de documento: Article