Your browser doesn't support javascript.
loading
Validation and Search of the Ideal Cut-Off of the Sysmex UF-1000i® Flow Cytometer for the Diagnosis of Urinary Tract Infection in a Tertiary Hospital in Spain.
Millán-Lou, María I; García-Lechuz, Juan M; Ruiz-Andrés, María A; López, Concepción; Aldea, María J; Revillo, María J; Rezusta, Antonio.
Afiliação
  • Millán-Lou MI; Servicio de Microbiología, Hospital Universitario Miguel-Servet, Zaragoza, Spain.
  • García-Lechuz JM; Instituto de Investigación Sanitaria Aragón (IIS Aragón), Zaragoza, Spain.
  • Ruiz-Andrés MA; Servicio de Microbiología, Hospital Universitario Miguel-Servet, Zaragoza, Spain.
  • López C; Servicio de Microbiología, Hospital Universitario Miguel-Servet, Zaragoza, Spain.
  • Aldea MJ; Servicio de Microbiología, Hospital Universitario Miguel-Servet, Zaragoza, Spain.
  • Revillo MJ; Servicio de Microbiología, Hospital Universitario Miguel-Servet, Zaragoza, Spain.
  • Rezusta A; Servicio de Microbiología, Hospital Universitario Miguel-Servet, Zaragoza, Spain.
Front Med (Lausanne) ; 5: 92, 2018.
Article em En | MEDLINE | ID: mdl-29686988
Urinary tract infections (UTI) are one of the most prevalent infections. A rapid and reliable screening method is useful to screen out negative samples. The objective of this study was to validate the Sysmex flow cytometer UF-1000i by evaluating its accuracy, linearity and carry-over; and define an optimal cut-off value to be used in routine practice in our hospital. For the validation of the UF-1000i cytometer, precision, linearity and carry-over were studied in samples with different counts of bacteria, leukocytes and erythrocytes. Between March and June 2016, urine samples were tested in the Clinical Microbiology Laboratory at University Miguel Servet Hospital, in Spain. Samples were analyzed with the Sysmex UF-1000i cytometer, and cultured. Growth of ≥105 CFUs/mL was considered positive. The validation study reveals that the precision in all the variables is acceptable; that there is a good linearity in the dilutions performed, obtaining values almost identical to those theoretically expected; and for the carry-over has practically null values. A total of 1,220 urine specimens were included, of which 213 (17.4%) were culture positive. The optimal cut-off point of the bacteria-leukocyte combination was 138.8 bacteria or 119.8 leukocytes with an S and E of 95.3 and 70.4%, respectively. The UF-1000i cytometer is a valuable method to screen urine samples to effectively rule out UTI and, may contribute to the reduction of unnecessary urine cultures.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: Front Med (Lausanne) Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Espanha País de publicação: Suíça

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: Front Med (Lausanne) Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Espanha País de publicação: Suíça