Diagnostic ability of ganglion cell complex thickness to detect glaucoma in high myopia eyes by Fourier domain optical coherence tomography.
Int J Ophthalmol
; 11(5): 791-796, 2018.
Article
em En
| MEDLINE
| ID: mdl-29862177
AIM: To evaluate the ability of macular ganglion cell complex (GCC) thickness using Fourier domain optical coherence tomography (FD-OCT) to detect glaucoma in highly myopic eyes. METHODS: Cross-sectional study. A total of 114 participants, consecutively were enrolled. Macular GCC thickness and peripapillary retinal nerve fiber layer (RNFL) thickness were obtained with RTVue FD-OCT. Receiver operating characteristics curves were constructed for each measurement parameter, and areas under the curves (AUCs) were compared. RESULTS: Both the average GCC and average RNFL thickness showed negative correlations with axial length (rGCC=-0.404, P=0.001; rRNFL=-0.561, P<0.001). The largest AUCs from GCC, and RNFL parameters were 0.968 [global loss volume (GLV)], and 0.855 (average RNFL), respectively. GLV was significantly better for detecting high myopic glaucoma than average RNFL (P<0.001). CONCLUSION: Macular GCC thickness has higher diagnostic power than peripapillary RNFL thickness to discriminate glaucoma patients from non-glaucoma subjects in high myopia.